27
(4) When the desired unit is displayed, press the
→
O/T
←
key to
register it.
(5) To return to the mass display, press and hold the POWER/BRK key.
Release of the Measurement Unit
Perform operations (1) to (3) above to display the units. When the currently
registered unit with the stability symbol is displayed, press the
→
O/T
←
key
to release registration of that unit.
Selection of Units
Press the UNIT key during mass display to switch between the registered
units.
If reference values are not set, as with % and PCS,
" --- "
is displayed
instead of a numeric value.
F
See Section
10.1 “Percentage”
or Section
10.2 “Number of Pieces”
for
more information on setting reference values
.
10.1
Percentage
Procedure
This balance is capable of displaying mass values in the % unit by setting a
reference sample as 100%.
(1) Register the % unit.
F
See Section
10. “REGISTRATION, RELEASE, AND
SELECTION OF MEASUREMENT UNITS.”
If % unit is already
registered, it is not necessary to register it again.
(2) Place the weighing vessel on the pan and press the
→
O/T
←
key.
(3) Add the reference sample.
(4) From the mass display, press the UNIT key to select the % unit.
(5) Press and hold the CAL/MENU key until
"SEt 100%"
is displayed.
(6) Confirm that the stability symbol is lit and press the
→
O/T
←
key.
(7)
"SEt"
is displayed for several seconds and then the % unit is
displayed.
The minimum display value varies depending on the mass of the reference
sample. (REF.)
The following reference sample mass is the count value when the minimum
display value is regarded as 1.