39
entry.
"HH:MM"
(HH: hour, MM: min) is displayed and the
leftmost digit is blinking.
(3) Change the blinking digit by pressing the UNIT key to increase the
numeric value by one. The hour can be set from 00 to 23 and the
minutes and seconds from 00 to 59. By setting all of the digits to
"_" the time setting is cancelled. Press the PRINT key to confirm
the current digit and move the blinking cursor to the next digit.
Press the POWER/BRK key to complete the changes. If an illogical
time is entered,
"Err 20"
is displayed.
(3) When
"tCAL t*"
is displayed, press and hold the POWER/BRK key
to return to mass display.
23. SOLVENT SPECIFIC GRAVITY FOR SOLID SPECIFIC GRAVITY
MEASUREMENT
Procedure
(1) From the mass display, press the CAL/MENU key several times.
When
"SEttinG"
is displayed, press the
→
O/T
←
key.
(2) Press the CAL/MENU key several times. When
"LSG SEt"
is
displayed, press the
→
O/T
←
key. The [MENU] and # symbols are
displayed to indicate numeric value entry.
"SG*.****"
is displayed
with the digit next to the G blinking.
(3) Change the blinking digit by pressing the UNIT key. Each time the
UNIT key is pressed, the value changes from 1
→
2
→
• • •
→
8
→
9
→
0
→
1
→
• • •. Press the PRINT key to confirm that value
and move the blinking cursor to the next digit.
(4) Press the
→
O/T
←
key to confirm the numeric value of the solvent
specific gravity for solid specific gravity measurement or press the
POWER/BRK key to return to
"LSG SEt"
without changing the
value.
(5) When
"LSG SEt"
is displayed, press and hold the POWER/BRK
key to return to mass display.