Error Message
Items identified by [S] may require factory service. Please contact your CAS
corporation representative if these error messages are displayed.
Error
Message
Description
Corrective Action
CAL E2
Zero point deviates largely
during span calibration.
Empty the pan.
CAL E3
Large deviation during
PCAL.
Use the correct
calibration mass.
CAL E4
Sensitivity deviates largely
during span calibration.
Use the correct
calibration mass.
CHE x
The balance stops at this
display.
→
[S]
Err 0x
The internal parts of the
balance require service.
→
[S]
Err 20
An incorrect value has been
entered.
Review the entered
numeric value.
Err21
Required conditions and
numeric values are not set.
Check the analog display
mode.
Err24
The balance cannot store
settings due to abnormal
power supply voltage.
Check the power supply
voltage.