Technical Data
7-9
Table 14: Ultra-low Variable Energy (150 VE) Powerheart G3 Waveform
Phase 1
Phase 2
Patient’s
Impedance
(Ohms)
Voltage*
(Volts)
Duration*
(MS)
Voltage*
(Volts)
Duration*
(MS)
Energy**
(Joules)
25
1393
3.3
743
3.2
145-196
50
1420
4.5
909
3.2
128-173
75
1430
5.8
973
3.2
116-156
100
1434
7.0
1007
3.2
108-146
125
1437
8.3
1027
3.2
102-138
150
1439
9.5
1040
3.2
98-132
175
1441
10.8
1049
3.2
95-128
Table 15: Low Variable Energy (200 VE) Powerheart G3 Waveform
Phase 1
Phase 2
Patient’s
Impedance
(Ohms)
Voltage*
(Volts)
Duration*
(MS)
Voltage*
(Volts)
Duration*
(MS)
Energy**
(Joules)
25
1609
3.3
858
3.2
193-260
50
1640
4.5
1050
3.2
170-230
75
1651
5.8
1124
3.2
155-209
100
1656
7.0
1163
3.2
144-194
125
1660
8.3
1186
3.2
136-184
150
1662
9.5
1201
3.2
131-176
175
1663
10.8
1212
3.2
126-170