Version 112-2025-109 B
Product Information and Safety
1-17
Recommended separation distances between portable and
mobile RF communications equipment and the AED
The AED is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the
AED can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications
equipment (transmitters) and the AED as recommended below, according to
the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
outside ISM bands
d = 1.2
150 kHz to 80 MHz
in ISM bands
d = 1.2
80 MHz to 800
MHz
d = 1.2
800 MHz to 2.5
GHz
d = 2.3
0.01
0.12
0.12
0.12
0.23
0.1
0.38
0.38
0.38
0.73
1
1.2
1.2
1.2
2.3
10
3.8
3.8
3.8
7.3
100
12
12
12
23
P
P
P
P
Содержание Powerheart AED G3 Plus 9390A
Страница 1: ...OPERATOR AND SERVICE MANUAL Version 112 2025 109 B Revised February 4 2010 www AmericanAED com...
Страница 6: ...Table of Contents Version 112 2025 109 B vi...
Страница 10: ...112 2025 109 B Limited Warranty x...
Страница 60: ...Step 6 Post Rescue Version 112 2025 109 B 4 8...
Страница 70: ...Frequently Asked Questions Version 112 2025 109 B 6 8...