Parameters
Version 112-2025-109 B
7-6
Environmental Conditions
Free Fall Drop: IEC 60068-2-32 (1975) AM 2
(1990), 1 meter
Bump: IEC 60068-2-29 (1987), 40g and 6000
bumps
Vibration (Random): IEC 60068-2-64 (1993): 10Hz
– 2KHz, 0.005 – 0.0012 g
2
/Hz
Vibration (Sine): IEC 60068-2-6 (1995): 10Hz –
60Hz, 0.15 mm and 60Hz – 150Hz, 2g
Enclosure Protection: IEC 60529 (2001), IP24
Shipping and
Transportation Conditions
ISTA Procedure 2A
RHYTHMx ECG Analysis
Performance
The AED RHYTHMx ECG Analysis system
analyzes the patient’s ECG and advises you when
the AED detects a shockable or non-shockable
rhythm.
This system makes it possible for a person, with
no training in the interpretation of ECG rhythms,
to offer defibrillation therapy to victims of sudden
cardiac arrest.
Table 13: Parameters (continued)
Parameter
Detail