Version 112-2025-109 B
Getting Started
3-17
“Press Pads Firmly to
Patient’s Bare Skin.”
PRESS PADS FIRMLY
TO BARE SKIN
Prompts when better pad connectivity
to the patient’s skin is required because
impedance is too high.
Table 11: Other Prompts
Voice Prompt
Text Display
Situation
“Battery Low”
BATTERY LOW
Occurs once when the battery voltage
becomes low, although a rescue can
continue for approximately 9 more
shocks. When the battery is too low to
do a rescue, the following will occur:
1
“BATTERY LOW” will show on the LCD
2
STATUS INDICATOR will turn RED
3
AED will BEEP once every 30
seconds
You must replace the battery before
continuing with the rescue. If
completely depleted, all AED activity
will terminate.
“Analysis Interrupted.
Stop Patient Motion.”
ANALYSIS INTERRUPTED
STOP PATIENT MOTION
When the AED detects ECG noise
artifact, stop moving or touching the
patient. Remove other electronic
devices within a 5 meter radius.
“Open Lid to Continue
Rescue.”
OPEN LID
TO CONTINUE RESCUE
When the lid is inadvertently closed
during a rescue, this prompt will repeat
for 15 seconds.
“Rhythm Changed.
Shock Cancelled.”
RHYTHM CHANGED
SHOCK CANCELLED
When the device is prepared to shock
then detects a change in rhythm and
therefore cancels the shock.
“Remove Cable to
Continue Rescue.”
REMOVE CABLE
TO CONTINUE RESCUE
When a serial communication cable is
connected to the AED during a rescue,
the phrase repeats until the cable is
disconnected.
Table 10: Pad Issues
Voice Prompt
Text Display
Situation
Содержание Powerheart AED G3 Plus 9390A
Страница 1: ...OPERATOR AND SERVICE MANUAL Version 112 2025 109 B Revised February 4 2010 www AmericanAED com...
Страница 6: ...Table of Contents Version 112 2025 109 B vi...
Страница 10: ...112 2025 109 B Limited Warranty x...
Страница 60: ...Step 6 Post Rescue Version 112 2025 109 B 4 8...
Страница 70: ...Frequently Asked Questions Version 112 2025 109 B 6 8...