Version 112-2025-109 B
Getting Started
3-13
The following table lists the voice and text prompts and a description of
when the prompts are issued.
Table 4: Initial Instructions
Voice Prompt
Text Display
Situation
“Stay Calm. Follow These
Voice Instructions. Make
Sure 911 is Called Now!”
CALL 911!
Plays after lid opening self test, default
ON.
“Stay Calm. Follow These
Voice Instructions. Make
Sure Emergency Services
are Called Now!”
CALL EMERGENCY
SERVICES NOW!
Medical Director may use MDLink to
select this prompt instead of “CALL
911!”.
MDLink also allows the emergency
services and 911prompt to be disabled.
Table 5: Preparation
Voice Prompt
Text Display
Situation
“Begin by Exposing
Patient’s Bare Chest and
Torso. Remove or Cut
Clothing if Needed.”
BARE PATIENT’S TORSO
REMOVE CLOTHING
Prompts the rescuer to remove patient
clothing.
“When Patient’s Chest and
Torso are Exposed,
Remove Square Foil
Package from Lid of AED.”
WHEN CHEST IS BARE
REMOVE FOIL PACKAGE
Prompts the rescuer to remove the
pads from AED lid.
“Tear Open Foil Package
Across Dotted Line and
Remove Pads.”
TEAR OPEN PACKAGE
REMOVE PADS
Prompts the rescuer to open the pad
package and remove pads.
“Next, Separate One of
the White Pads
Completely from Blue
Plastic Liner. Begin Peeling
from the Tabbed Corner.”
PEEL ONE PAD FROM
BLUE PLASTIC LINER
Repeats every 3 seconds until the pads
are separated. If a pad has been peeled
before the prompt starts, this prompt
will be skipped. This prompt will be
interrupted when pad is peeled.
Содержание Powerheart AED G3 Plus 9390A
Страница 1: ...OPERATOR AND SERVICE MANUAL Version 112 2025 109 B Revised February 4 2010 www AmericanAED com...
Страница 6: ...Table of Contents Version 112 2025 109 B vi...
Страница 10: ...112 2025 109 B Limited Warranty x...
Страница 60: ...Step 6 Post Rescue Version 112 2025 109 B 4 8...
Страница 70: ...Frequently Asked Questions Version 112 2025 109 B 6 8...