Rescue Protocol
Version 112-2025-109 B
2-4
SVT Rate
All rhythms with rates between the Detection Rate and SVT Rate will be
screened through a number of SVT Discriminators to classify them into VF/
VT or SVT. Rhythms classified as SVT between the two set rates are not
shockable. All SVT rhythms above the rates will be classified as shockable.
The SVT Rate must be greater than the Detection Rate and is selectable
between 160 and 300 bpm or, “NO THERAPY FOR SVT” can be selected
via MDLink Software by the Medical Director.
Rescue Protocol
The AED rescue protocol is consistent with the guidelines recommended by
the American Heart Association (
Guidelines 2005 for Cardiopulmonary
Resuscitation and Emergency Cardiac Care
American Heart Association;
Circulation vol 112, Issue 24 Suppl. Dec. 13, 2005) and the International
Liaison Committee on Resuscitation (ILCOR)).
Upon detecting a shockable cardiac rhythm, the AED advises the operator to
press the SHOCK button (9390E only) to deliver a defibrillation shock
followed by performing 2 minutes of CPR.
For the Powerheart AED G3 Automatic, upon detecting a shockable rhythm,
the AED will automatically deliver defibrillation shocks followed by
performing 2 minutes of CPR.
Note:
In alignment with the 2005 Guidelines, the default setting for the
CPR time has been set to allow for 5 cycles of 30 compressions and 2
breaths. Increasing or decreasing the CPR time setting may increase or
decrease the number of actual cycles allowed during the CPR time out
period.
STAR Biphasic Waveform
The STAR
®
Biphasic Waveform is designed to measure the patient’s
impedance and deliver a customized shock. This allows the delivery of an
optimized energy level to each patient. The energy levels for the Powerheart
G3 AED are available in three different defibrillation shock levels.
The Ultra-Low Energy (150 VE), Low Energy (200 VE), and High Energy
(300 VE) shocks are variable energy. The actual energy is determined by the
patient’s impedance. See Table 2 on page 2-5, Table 14 on page 7-9, Table 15
on page 7-9, and Table 16 on page 7-10 for additional information.
Содержание Powerheart AED G3 Plus 9390A
Страница 1: ...OPERATOR AND SERVICE MANUAL Version 112 2025 109 B Revised February 4 2010 www AmericanAED com...
Страница 6: ...Table of Contents Version 112 2025 109 B vi...
Страница 10: ...112 2025 109 B Limited Warranty x...
Страница 60: ...Step 6 Post Rescue Version 112 2025 109 B 4 8...
Страница 70: ...Frequently Asked Questions Version 112 2025 109 B 6 8...