AED Indicators
Version 112-2025-109 B
3-8
!
Caution. Possible Improper AED Performance.
Using pads that are damaged or expired may result in improper AED
performance.
Directions for Use
Pads are for short term use only. Do not open until ready to use.
!
Caution. Equipment Damage.
Do not pull on the lead wire to separate the pads from the blue liner. Follow
directions on the pad packaging:
1
Ensure the skin site is clean and dry.
2
Separate one pad from blue liner.
3
Place one pad on bare skin in either location.
4
Peel and place remaining pad in opposite location.
AED Indicators
The following indicators are located on the AED.
RescueReady Status Indicator
The STATUS INDICATOR is located on the Powerheart G3 AED handle.
When this indicator is GREEN, the AED is RescueReady. This means the
AED self-tests have verified the following:
◆
Battery has an adequate charge
◆
Pads are properly connected to the AED and functioning
◆
Integrity of the internal circuitry is good
Содержание Powerheart AED G3 Plus 9390A
Страница 1: ...OPERATOR AND SERVICE MANUAL Version 112 2025 109 B Revised February 4 2010 www AmericanAED com...
Страница 6: ...Table of Contents Version 112 2025 109 B vi...
Страница 10: ...112 2025 109 B Limited Warranty x...
Страница 60: ...Step 6 Post Rescue Version 112 2025 109 B 4 8...
Страница 70: ...Frequently Asked Questions Version 112 2025 109 B 6 8...