Version 112-2025-109 B
Getting Started
3-7
The pads have a limited shelf life and should not be used beyond the
expiration date. Keep a fresh, unopened pair of pads plugged into the AED at
all times. Refer to the pad package label for operation temperatures.
An audible and visual alert will indicate after the self-test if the pads are
missing, unplugged, or damaged.
Pad Installation
To install the pads:
1
Open the lid of the AED.
2
Place the pad package into the lid so that the expiration label is visible
through the clear window on the lid. The expiration date of the pads will
then be readable without opening the lid of the AED.
3
Match the color of the connectors (red to red), then plug the pad
connector into the AED case as shown in the drawing. Once the pad
connector is plugged into AED, the PAD indicator should extinguish.
4
Tuck the excess cable length in the bottom holder as shown in the
drawing. With the pad package completely secured to the AED lid, close
the lid.
5
Make sure the expiration date is visible through the clear window of the
lid and check to make sure that the STATUS INDICATOR is GREEN. If
the pads are not installed properly, the STATUS INDICATOR will be
RED. Call Customer Service for assistance.
!
Caution. Use only Cardiac Science Approved Equipment.
Using batteries, pads, cables, or optional equipment other than those approved
by Cardiac Science may cause the AED to function improperly during a rescue.
Содержание Powerheart AED G3 Plus 9390A
Страница 1: ...OPERATOR AND SERVICE MANUAL Version 112 2025 109 B Revised February 4 2010 www AmericanAED com...
Страница 6: ...Table of Contents Version 112 2025 109 B vi...
Страница 10: ...112 2025 109 B Limited Warranty x...
Страница 60: ...Step 6 Post Rescue Version 112 2025 109 B 4 8...
Страница 70: ...Frequently Asked Questions Version 112 2025 109 B 6 8...