Indicator Troubleshooting Table
Version 112-2025-109 B
6-2
Self-tests will be initiated upon opening the lid and again upon closing the
lid. If the self-test detects an error, the Status Indicator will remain RED.
Upon closing the lid, an audible alert will be issued. The Diagnostic Panel
under the lid will indicate the source of the problem according to the
Indicator Troubleshooting Guide Table on the next page.
Indicator Troubleshooting Table
The following is a troubleshooting table for the AED indicators.
Table 12: Indicator Troubleshooting Table
View
Symptom
Solution
Red SERVICE indicator (LED) is
lit.
Maintenance by authorized
service personnel is required.
Call Cardiac Science Customer
Service (see page 4) or your local
Cardiac Science distributor.
Red Pads indicator (LED) is lit.
Connect the pads or replace
with a new pair.
The last battery indicator (LED) is
red.
The battery is low. Replace with
a new battery.
STATUS INDICATOR is RED, and
no other indicators on the
diagnostic panel are lit.
The battery power is completely
depleted. Replace with a new
battery. If STATUS INDICATOR
remains RED call Cardiac
Science Customer Service or
your local Cardiac Science
distributor.