Getting Started
3-3
Standby Mode:
When the battery is installed, but the lid is closed. In this
mode the AED is not being used in a rescue. The device will conduct its
routine self-tests to ensure proper operation.
Storage Mode:
When the battery is removed, such as during shipping or
transport. With the battery removed, the AED is unable to perform self-tests
or rescues.
Environmental Operating and Standby Conditions
!
Caution. Temperature Extremes.
Exposing the AED to extreme environmental conditions outside of its operating
parameters may compromise the ability of the AED to function properly. The
RescueReady
®
daily self-test verifies the impact of extreme environmental
conditions on the AED. If the daily self-test determines environmental conditions
outside of the AED’s operating parameters, a “Service Required” alert will be
issued to prompt the user to move the AED to environmental conditions within
the acceptable operating parameters at once.
Shipping and Transport Conditions
(For up to 1 week)
IntelliSense Battery
IntelliSense batteries contain an integrated memory chip that automatically
stores important usage information, enabling the battery to maintain a
complete history of its operating life. The actual battery history can be
reviewed using the RescueLink software.
Содержание Powerheart AED G3 Plus 9390A
Страница 1: ...OPERATOR AND SERVICE MANUAL Version 112 2025 109 B Revised February 4 2010 www AmericanAED com...
Страница 6: ...Table of Contents Version 112 2025 109 B vi...
Страница 10: ...112 2025 109 B Limited Warranty x...
Страница 60: ...Step 6 Post Rescue Version 112 2025 109 B 4 8...
Страница 70: ...Frequently Asked Questions Version 112 2025 109 B 6 8...