Version 112-2025-109 B
Getting Started
3-5
!
Caution. Possible Improper AED Performance.
Using pads that are damaged or expired may result in improper AED
performance.
Note:
Battery part number 9146 is only for use with the Powerheart AED
G3 and Powerheart AED G3 Automatic.
Battery Operating Life
The battery operating life depends on the type of battery, actual usage and
environmental factors.
The following table represents the expected life of the Powerheart G3 AED
when used in Standby Mode.
Battery Shelf Life
The batteries have a shelf life of five years. Shelf life is defined as the length
of time a battery can be stored, prior to installation into AED, without
degrading its performance.
Note:
Storing the battery outside its specific range (0-50°C) will decrease
battery life.
Battery Installation
To install the battery:
Table 3: Normal Battery Operating Life
Model
Estimated Shelf
Life (from date
of manufacture)
Full Operational
Replacement
Guarantee (from date
of installation)
Typical Shocks
9146 Lithium
(p/n: 9146-001)
5 Years
4 Years
up to 290