Version 112-2025-109 B
Product Information and Safety
1-13
Guidance and manufacturer’s
declaration—electromagnetic immunity
The AED is intended for use in the electromagnetic environment specified
below. The customer or the user of the AED should assure that it is used in
such an environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment—
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
Not applicable
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common mode
Not applicable