Voice Prompts and Text Display
Version 112-2025-109 B
3-18
“Communications
Mode”
COMMUNICATIONS MODE When the lid is open and the serial
communication cable is plugged into
the AED.
“Service Required”
SERVICE REQUIRED
Occurs after the self-tests determine
that the AED is not functioning
properly. The prompt “SERVICE
REQUIRED” will be heard when the lid
is opened. The red SERVICE indicator
will illuminate and “SERVICE
REQUIRED” will repeat until you close
the lid. After closing the lid, an alarm
beep will be heard until the battery is
removed or becomes completely
depleted.
Table 11: Other Prompts (continued)
Voice Prompt
Text Display
Situation
Содержание Powerheart AED G3 Plus 9390A
Страница 1: ...OPERATOR AND SERVICE MANUAL Version 112 2025 109 B Revised February 4 2010 www AmericanAED com...
Страница 6: ...Table of Contents Version 112 2025 109 B vi...
Страница 10: ...112 2025 109 B Limited Warranty x...
Страница 60: ...Step 6 Post Rescue Version 112 2025 109 B 4 8...
Страница 70: ...Frequently Asked Questions Version 112 2025 109 B 6 8...