Step 3: ECG Analysis
When the pads are placed, the AED will prompt “Do Not Touch Patient.
Analyzing Heart Rhythm. Please Wait.” The AED will begin to analyze the
cardiac rhythm of the patient.
If the pads become disconnected from the AED, the prompt “Make sure pad
connector is plugged into AED” will be heard. If this occurs, check to be sure
the connector is properly plugged into the AED. If the pads are not properly
placed or become loose during the rescue, the voice prompt “Press pads
firmly to patient’s bare skin” will be heard. When this occurs, ensure that
pads are firmly placed on clean, dry skin.
If noise is detected during analysis, the AED will warn you with the prompt
“Analysis Interrupted. Stop Patient Motion” and restart the analysis. This
usually occurs if the patient is excessively jostled or there is a strong
electromagnetic emitting electronic device nearby (within 5 meters). Remove
the electronic device or stop the excessive motion if you hear this prompt.
Step 4: Shock Delivery
If a shock is advised, the voice prompt will say, “Preparing Shock. Move
away from the patient.”
You should ensure that you and any bystanders are not touching the patient.
For the Powerheart AED G3:
When the AED is ready to deliver a
defibrillation shock, the Shock button will flash and the prompt “Press Red
Flashing Button to Deliver Shock” will be heard. Make sure no one is
touching the patient and press the Shock button to deliver a defibrillation
shock. If you do not press the Shock button within 30 seconds of hearing the
prompt, the AED will disarm and prompt you to start CPR.
For the Powerheart AED G3 Automatic:
When the AED is ready to deliver a
shock, the voice prompt, “Shock Will be Delivered in Three, Two, One” then
deliver a shock. Make sure no one is touching the patient.
Both models:
After the AED delivers the defibrillation shock, the voice
prompt will say “Shock Delivered,” and prompt you to start CPR.
When the AED is charged, it continues to analyze the patient's heart rhythm.
If the rhythm changes and a shock is no longer needed, the AED will issue
Содержание Powerheart AED G3 Plus 9390A
Страница 1: ...OPERATOR AND SERVICE MANUAL Version 112 2025 109 B Revised February 4 2010 www AmericanAED com...
Страница 6: ...Table of Contents Version 112 2025 109 B vi...
Страница 10: ...112 2025 109 B Limited Warranty x...
Страница 60: ...Step 6 Post Rescue Version 112 2025 109 B 4 8...
Страница 70: ...Frequently Asked Questions Version 112 2025 109 B 6 8...