86
Cycle Operation
Startup Phase
During start-up phase, the software monitors certain sensors. If any of the monitored sensors are not satisfied during start-
up, an error message is displayed and the process is halted. To cancel the start-up phase, press the STOP key. To resume,
resolve the error according to the error message displayed, press the STOP key, and then press the Start key. Please refer to
the maintenance and troubleshooting section for appropriate instructions to resolve error messages.
If the leak tester is enabled, a 40 second sheath test is activated during the start-up phase. During that test the endoscope is
pressurized to 160mmhg for 20 seconds then the pressure is monitored for the remaining 20 seconds. If the pressure drops
below 50mmhg, a “Sheath Fail” error message is displayed and the process is halted. To cancel the start-up phase, press
the STOP key. To resume, resolve the issue according to the error message displayed, press the STOP key, and then press
the START key.
The cycle then proceeds with a default Flush Phase or an optional Wash Phase.
Flush Phase
The flush phase introduces fresh water through scope channels for a minimum of 30 seconds as well as into the basin via
the basin port. Detergent can be injected into the channels prior to the 30 second minimum fresh water flush via SETUP
5. The default detergent injection time is 3 seconds (~9 ml). Please refer to the table in the Programming the Reprocessor
Section for the appropriate dilution and timing settings. The flush time can also be increased above 30 seconds via SETUP
5. Refer to the Programming the Reprocessor Section.
Note: During this phase, the incoming water temperature is monitored to
ensure that the water temperature is within the proper range for disinfection. If the water temperature is below the minimum required
for disinfection, water will continue to flow into the basin until either the minimum incoming water temperature is met or the step
times out.
This flush phase concludes by purging the scope channels with air and emptying the basin of water.
Содержание MEDIVATORS DSD EDGE
Страница 1: ...1 Service Manual System Version 5 65...
Страница 23: ...23 Figure 5 Function Keys...
Страница 27: ...27 Figure 6 Set Date Screen...
Страница 29: ...29 Figure 7 Set Time Screens...
Страница 31: ...31 Figure 8 Software Version Screen Flash Code Version or Software version Micro controller Code Version...
Страница 33: ...33 Figure 9 Display Log Screen and sample entry...
Страница 42: ...42 Figure 14 Custom Program Setup Screens P1 Detergent Inject Set Minutes 00 00...
Страница 43: ...43 Figure 15 Custom Program Setup Screen...
Страница 51: ...51 Figure 22 Set Delayed Startup Screen...
Страница 54: ...54...
Страница 96: ...96 Notes...
Страница 101: ...101 Figure 4 Depress locking ring Figure 5 Insert tube Figure 6 Locking ring out...
Страница 107: ...107 Figure 14 Install the new water filter Figure 15 Reconnect the water filter Figure 16 Filter installed...
Страница 109: ...109 Figure 17 LCD Display Figure 18 Drain Purge Line Figure 19 Replace Pre filters 1 MICRON 0 45 MICRON...
Страница 111: ...111 Figure 20 Printer compartment Figure 21 Advance paper Figure 22 Feed paper through slot Figure 23 Install new roll...
Страница 115: ...115 Figure 28 Filter locations Figure 29 Disconnect the filter Figure 30 Verify inlet side...
Страница 119: ...119 Figure 33 Fuse locations...
Страница 121: ...121 Figure 34 Connect pressure gauge Figure 35 Bleed some air...
Страница 140: ...140 DSD EDGE W O RECIRC W ADDITIONAL ISOLATION VALVE W CHAMBER FLOW SWITCH HYDRAULIC SCHEMATIC REVISION 02...
Страница 141: ...141...
Страница 142: ...142...
Страница 143: ...143...
Страница 149: ...149...
Страница 150: ...150...
Страница 151: ...151...