Programming the Reprocessor
INPUT PROGRAM
Custom programs allow the operator to change the cycle parameter settings, or to setup custom reprocessing
protocols. A maximum of nine custom programs can be pre-set. Refer to the disinfection cycle chart in the appendix
for range settings.
Note: Depending on selections, some of the following screens will
not be displayed.
Note: To deactivate a cycle, enter “0” for the time setting, then press
the ENTER button.
1. Press the SETUP button.
•
Enter 5 on the keypad, then press the ENTER button.
2. The “Program 1” screen is displayed. Enter the program digit (1-9) on the numeric keypad, then press the
ENTER button.
3. The “Flush” screen is displayed. Enter the desired detergent flush time.
•
Enter two digits for the minutes, then press ENTER.
•
Enter two digits for the seconds, then press ENTER.
4. The “Soak” screen is displayed. Enter the desired soak time.
•
Enter two digits for the minutes, then press ENTER.
•
Enter two digits for the seconds, then press ENTER.
5. The “Soak Rinse” screen is displayed. Enter the desired soak rinse time.
•
Enter two digits for the minutes, then press ENTER.
•
Enter two digits for the seconds, then press ENTER.
Содержание MEDIVATORS DSD EDGE
Страница 1: ...1 Service Manual System Version 5 65...
Страница 23: ...23 Figure 5 Function Keys...
Страница 27: ...27 Figure 6 Set Date Screen...
Страница 29: ...29 Figure 7 Set Time Screens...
Страница 31: ...31 Figure 8 Software Version Screen Flash Code Version or Software version Micro controller Code Version...
Страница 33: ...33 Figure 9 Display Log Screen and sample entry...
Страница 42: ...42 Figure 14 Custom Program Setup Screens P1 Detergent Inject Set Minutes 00 00...
Страница 43: ...43 Figure 15 Custom Program Setup Screen...
Страница 51: ...51 Figure 22 Set Delayed Startup Screen...
Страница 54: ...54...
Страница 96: ...96 Notes...
Страница 101: ...101 Figure 4 Depress locking ring Figure 5 Insert tube Figure 6 Locking ring out...
Страница 107: ...107 Figure 14 Install the new water filter Figure 15 Reconnect the water filter Figure 16 Filter installed...
Страница 109: ...109 Figure 17 LCD Display Figure 18 Drain Purge Line Figure 19 Replace Pre filters 1 MICRON 0 45 MICRON...
Страница 111: ...111 Figure 20 Printer compartment Figure 21 Advance paper Figure 22 Feed paper through slot Figure 23 Install new roll...
Страница 115: ...115 Figure 28 Filter locations Figure 29 Disconnect the filter Figure 30 Verify inlet side...
Страница 119: ...119 Figure 33 Fuse locations...
Страница 121: ...121 Figure 34 Connect pressure gauge Figure 35 Bleed some air...
Страница 140: ...140 DSD EDGE W O RECIRC W ADDITIONAL ISOLATION VALVE W CHAMBER FLOW SWITCH HYDRAULIC SCHEMATIC REVISION 02...
Страница 141: ...141...
Страница 142: ...142...
Страница 143: ...143...
Страница 149: ...149...
Страница 150: ...150...
Страница 151: ...151...