PRE-FILTER REPLACEMENT
Verify both station are idle before performing this procedure..
1. Close Incoming Water Line.
2. Open Internal Water Supply Valve.
• Press the SETUP button. Press 43 on the keypad, then press ENTER. This will open the incoming water
supply solenoid vlave providing a water supply to the on-board 0.2 micron filter housing.
NOTE: A counter on the LCD display will begin to count down from 250 to 0 in one second
increments. After 250 seconds, the incoming water supply solenoid valve will automatically
close if not manually closed sooner.
3. Drain Purge Line
• Direct the purge line from the top of the on-board 0.2 micron water filter housing into a container that holds
a minimum of 1 liter of liquid.
•
Slowly open the shut-off valve located on the purge line from the on-board 0.2 nmicron filter housing until
the air is purged from the filter housing and a steady stream of water flows from the purge lines.
• After closing the air bleed valve, tap the purge line to promote the draining of the residual water within the
purge line into the container.
• Remove the container and empty.
• Press the STOP button to close the incoming water supply solenoid valve and then press the STOP button
again to exit the SETUP menu.
4. Replace Pre-filters
• Use the fitler wrench, if necessary, to remove the pre-filter housings.
•
Wearing gloves, remove and discard the filters.
• Clean housings using soap and water and then rinse thoroughly.
•
Replace the corresponding fitler(s).
•
Replace the filter housing. You may need to moisten the O-ring located at the top of the filter housings with a
water-based lubricant. Make sure that the O-ring is at the top and in place in the top of the housing.
•
Hand tighten the filter housings; DO NOT use the filter wrench.
5. Open Incomig Water Line Slowly
6. Open Internal Water Supply Valve
• Follow the instructions in Step 2.
7. Drain Purge Line
• Follow the instructions in Step 3.
8. Check for Leaks
• If a leak is found, close the incoming water supply line and hand-tighten
the housing again.
• Open the water line again and check for leaks.
Содержание MEDIVATORS DSD EDGE
Страница 1: ...1 Service Manual System Version 5 65...
Страница 23: ...23 Figure 5 Function Keys...
Страница 27: ...27 Figure 6 Set Date Screen...
Страница 29: ...29 Figure 7 Set Time Screens...
Страница 31: ...31 Figure 8 Software Version Screen Flash Code Version or Software version Micro controller Code Version...
Страница 33: ...33 Figure 9 Display Log Screen and sample entry...
Страница 42: ...42 Figure 14 Custom Program Setup Screens P1 Detergent Inject Set Minutes 00 00...
Страница 43: ...43 Figure 15 Custom Program Setup Screen...
Страница 51: ...51 Figure 22 Set Delayed Startup Screen...
Страница 54: ...54...
Страница 96: ...96 Notes...
Страница 101: ...101 Figure 4 Depress locking ring Figure 5 Insert tube Figure 6 Locking ring out...
Страница 107: ...107 Figure 14 Install the new water filter Figure 15 Reconnect the water filter Figure 16 Filter installed...
Страница 109: ...109 Figure 17 LCD Display Figure 18 Drain Purge Line Figure 19 Replace Pre filters 1 MICRON 0 45 MICRON...
Страница 111: ...111 Figure 20 Printer compartment Figure 21 Advance paper Figure 22 Feed paper through slot Figure 23 Install new roll...
Страница 115: ...115 Figure 28 Filter locations Figure 29 Disconnect the filter Figure 30 Verify inlet side...
Страница 119: ...119 Figure 33 Fuse locations...
Страница 121: ...121 Figure 34 Connect pressure gauge Figure 35 Bleed some air...
Страница 140: ...140 DSD EDGE W O RECIRC W ADDITIONAL ISOLATION VALVE W CHAMBER FLOW SWITCH HYDRAULIC SCHEMATIC REVISION 02...
Страница 141: ...141...
Страница 142: ...142...
Страница 143: ...143...
Страница 149: ...149...
Страница 150: ...150...
Страница 151: ...151...