132
Set D Pulse Secs .................... Diagnostic 64: prompts the user to enter the disinfectant pulse time in seconds.
Set D Top Secs ........................ Diagnostic 67: prompts the user to enter the disinfectant top-off time in seconds.
Set Dis. Hold ........................... Diagnostic 69: prompts the user to enter the water line disinfectant hold time in hours
...............................................
and minutes.
Set Extra Air ............................ Diagnostic 65: prompts the user to enter the add extra air time in seconds.
Set F purge Secs .................... Diagnostic 63: prompts the user to enter the fluid purge time in seconds.
Set Idle State? ........................ Diagnostic 89: prompts the user to press the enter button to jump to the idle state.
Set P Drain Secs ..................... Diagnostic 66: prompts the user to enter the short rinse partial drain time in seconds.
Set P Fill Secs ........................ Diagnostic 66: prompts the user to enter the short rinse partial fill time in seconds.
Set R Drain Secs ..................... Diagnostic 61: prompts the user to enter the rinse drain time in seconds.
Set R Fill Secs ......................... Diagnostic 62: prompts the user to enter the rinse fill time in seconds.
Set R Rinse Secs ................... Diagnostic 68: prompts the user to enter the short rinse recirc time in seconds.
Set R Top Secs ....................... Diagnostic 67: prompts the user to enter the rinse top-off time in seconds.
Set Temp. to 10C .................... Diagnostic 31 and 32: prompts the user to set the thermistor to 10C.
Set Temp. to 30C .................... Diagnostic 31: prompts the user to set the thermistor to 30C.
Shth Sen Enable ..................... Diagnostic 45: prompts the user to enable / disable the sheath (leak) tester option.
Temp Option .......................... Diagnostic 88: prompts the user to enable / disable the temperature monitoring option.
Testing Sensors ..................... Diagnostic 94: indicates that the current sensor values are being displayed.
Time Limit .............................. Diagnostic 81: prompts the user to enter the programmable time limits.
Log Messages
A
dd Air ............................... Indicates that an add air cycle was run.
Add Air On .......................... Indicates that as Add Air cycle will be added at the end of the current cycle.
Add Air Off ......................... Indicates that as Add Air cycle will not be added at the end of the current cycle.
Air ..................................... Notes when the air portion of a cycle occurred.
Auto Dis ............................. Indicates that an auto water line disinfectant cycle was run.
Disinfect ............................ Notes when the disinfectant portion of a cycle occurred.
HLD Fail ............................. Log message indicating disinfectant efficacy has passed MRC test.
HLD Pass ........................... Log message indicating disinfectant efficacy has failed MRC test.
Low Alcohol ....................... Indicates that the alcohol level is low.
Machine ID ........................ The serial number of the DSD-201.
No Air Flow ........................ Indicates that no air flow was detected.
No Fluid Flo ........................ Indicates that no fluid flow was detected.
Power on ........................... Indicates that the device lost power during a cycle.
Resume ............................. Indicates when a cycle was resumed.
Rinse 1 .............................. Notes when the Rinse 1 portion of a cycle occurred.
Rinse 2 .............................. Notes when the Rinse 2 portion of a cycle occurred.
Rinse 3 .............................. Notes when the Rinse 3 portion of a cycle occurred.
S/N .................................... Indicates the machine serial number.
Sheath Fail ......................... Indicates that the sheet test failed.
Sheath Test ........................ Notes when the sheath test portion of a cycle occurred.
Содержание MEDIVATORS DSD EDGE
Страница 1: ...1 Service Manual System Version 5 65...
Страница 23: ...23 Figure 5 Function Keys...
Страница 27: ...27 Figure 6 Set Date Screen...
Страница 29: ...29 Figure 7 Set Time Screens...
Страница 31: ...31 Figure 8 Software Version Screen Flash Code Version or Software version Micro controller Code Version...
Страница 33: ...33 Figure 9 Display Log Screen and sample entry...
Страница 42: ...42 Figure 14 Custom Program Setup Screens P1 Detergent Inject Set Minutes 00 00...
Страница 43: ...43 Figure 15 Custom Program Setup Screen...
Страница 51: ...51 Figure 22 Set Delayed Startup Screen...
Страница 54: ...54...
Страница 96: ...96 Notes...
Страница 101: ...101 Figure 4 Depress locking ring Figure 5 Insert tube Figure 6 Locking ring out...
Страница 107: ...107 Figure 14 Install the new water filter Figure 15 Reconnect the water filter Figure 16 Filter installed...
Страница 109: ...109 Figure 17 LCD Display Figure 18 Drain Purge Line Figure 19 Replace Pre filters 1 MICRON 0 45 MICRON...
Страница 111: ...111 Figure 20 Printer compartment Figure 21 Advance paper Figure 22 Feed paper through slot Figure 23 Install new roll...
Страница 115: ...115 Figure 28 Filter locations Figure 29 Disconnect the filter Figure 30 Verify inlet side...
Страница 119: ...119 Figure 33 Fuse locations...
Страница 121: ...121 Figure 34 Connect pressure gauge Figure 35 Bleed some air...
Страница 140: ...140 DSD EDGE W O RECIRC W ADDITIONAL ISOLATION VALVE W CHAMBER FLOW SWITCH HYDRAULIC SCHEMATIC REVISION 02...
Страница 141: ...141...
Страница 142: ...142...
Страница 143: ...143...
Страница 149: ...149...
Страница 150: ...150...
Страница 151: ...151...