Water Line Disinfection
This function disinfects the water lines in the reprocessor. The current water quality standard for Washers/Disinfectors
in the United States and Canada recommends the use of potable (human consumption/drinking quality) water as the
supply water for use in washers/disinfectors. If you facility can supply potable quality water to the DSD EDGE
®
automated endoscope reprocessor, you will meet these requirements, so daily disinfection of the water lines and water
filter is not required, as long as the pre-filtration system is installed upstream of the reprocessor and is functioning
properly. Other countries have a more stringent water quality requirement under the standard ISO 15883-4:2008
(Washers-disinfectors Part 4). If you are required to meet this standard, chemical disinfection of the water treatment
(filtration) and delivery system must be performed on a weekly basis and whenever the 0.2 micron bacterial retentive
filter is replaced. This standard requires that all rinse water delivered to the washer/disinfector contain no more than 10
cfu (colony forming units) per 100 ml. Microbiological quality of even potable water varies considerably. If you are
unsure of the microbiological quality of your washer/disinfector supply water, you should have it tested according to
either local or internal guidelines. This procedure must also be performed after each water filter change and after any
service is performed on the water supply system.
Caution!
Ensure the restrictor adapter provided with the installation kit is connected in
the basin before performing this procedure.
1. Verify both stations are idle before performing this procedure. The default disinfection time is one and one-half
(1-1/2) hour.
2. Press the SETUP button.
• Enter 6 on the keypad, then press the ENTER button.
3. Press the START button.
• The LCD displays a reminder message to “Attach Restrictor”. Press the START button again after
verifying the restrictor is connected.
• The disinfectant will remain in the lines for the pre-programmed water line disinfection time.
!
Содержание MEDIVATORS DSD EDGE
Страница 1: ...1 Service Manual System Version 5 65...
Страница 23: ...23 Figure 5 Function Keys...
Страница 27: ...27 Figure 6 Set Date Screen...
Страница 29: ...29 Figure 7 Set Time Screens...
Страница 31: ...31 Figure 8 Software Version Screen Flash Code Version or Software version Micro controller Code Version...
Страница 33: ...33 Figure 9 Display Log Screen and sample entry...
Страница 42: ...42 Figure 14 Custom Program Setup Screens P1 Detergent Inject Set Minutes 00 00...
Страница 43: ...43 Figure 15 Custom Program Setup Screen...
Страница 51: ...51 Figure 22 Set Delayed Startup Screen...
Страница 54: ...54...
Страница 96: ...96 Notes...
Страница 101: ...101 Figure 4 Depress locking ring Figure 5 Insert tube Figure 6 Locking ring out...
Страница 107: ...107 Figure 14 Install the new water filter Figure 15 Reconnect the water filter Figure 16 Filter installed...
Страница 109: ...109 Figure 17 LCD Display Figure 18 Drain Purge Line Figure 19 Replace Pre filters 1 MICRON 0 45 MICRON...
Страница 111: ...111 Figure 20 Printer compartment Figure 21 Advance paper Figure 22 Feed paper through slot Figure 23 Install new roll...
Страница 115: ...115 Figure 28 Filter locations Figure 29 Disconnect the filter Figure 30 Verify inlet side...
Страница 119: ...119 Figure 33 Fuse locations...
Страница 121: ...121 Figure 34 Connect pressure gauge Figure 35 Bleed some air...
Страница 140: ...140 DSD EDGE W O RECIRC W ADDITIONAL ISOLATION VALVE W CHAMBER FLOW SWITCH HYDRAULIC SCHEMATIC REVISION 02...
Страница 141: ...141...
Страница 142: ...142...
Страница 143: ...143...
Страница 149: ...149...
Страница 150: ...150...
Страница 151: ...151...