![Cantel Medical MEDIVATORS DSD EDGE Скачать руководство пользователя страница 91](http://html1.mh-extra.com/html/cantel-medical/medivators-dsd-edge/medivators-dsd-edge_service-manual_3957444091.webp)
91
Leak Testing
Use the following procedure to leak test an endoscope. Leak test adaptors are available for PENTAX
®
, OLYMPUS
®
,
FUJIFILM
TM
and KARL STROZ
®
endoscopes.
Note: This automated test is not a substitute for the endoscope manufacturer’s manual leak
test. Follow the manufacturer’s instructions when performing a manual test.
1. Install the waterproof caps and leak tester adaptors on the endoscope following the endoscope manufacturer’s
instructions.
2. Load the endoscope into the reprocessor.
- Connect the leak tester hookup between the endoscope and the basin outlet.
3. Select the desired disinfection cycle. Press the START button on the reprocessor control panel.
4. The endoscope inflates for 20 seconds to 160mmHg (3psi). Endoscope pressure is monitored for another 20
seconds. -If the pressure decreases below the 50mmHg (1psi) reading within this period, the warning LED on the
reprocessor control panel blinks, the system activates an alarm and the message “Sheath Fail” is displayed. Press
the STOP button to end the cycle. The cycle is aborted.
- If no leak is detected, the disinfection cycle starts as normal. No indication is shown.
5. Pressure is maintained during the disinfection cycle to detect small leaks and prevent fluid ingression. If a small
leak is detected, the reprocessor will continue the cycle then alert the operator of any detected leaks at the end of
the cycle.
- Press the STOP button to acknowledge the warning.
6. The endoscope automatically deflates at the end of the cycle.
7. The log printout indicates if the leak tester option is disabled, or indicates any leak test failure.
Caution!
The leak test adaptor must be disconnected and removed from
the basin when not in use to avoid potential fluid ingression.
!
Содержание MEDIVATORS DSD EDGE
Страница 1: ...1 Service Manual System Version 5 65...
Страница 23: ...23 Figure 5 Function Keys...
Страница 27: ...27 Figure 6 Set Date Screen...
Страница 29: ...29 Figure 7 Set Time Screens...
Страница 31: ...31 Figure 8 Software Version Screen Flash Code Version or Software version Micro controller Code Version...
Страница 33: ...33 Figure 9 Display Log Screen and sample entry...
Страница 42: ...42 Figure 14 Custom Program Setup Screens P1 Detergent Inject Set Minutes 00 00...
Страница 43: ...43 Figure 15 Custom Program Setup Screen...
Страница 51: ...51 Figure 22 Set Delayed Startup Screen...
Страница 54: ...54...
Страница 96: ...96 Notes...
Страница 101: ...101 Figure 4 Depress locking ring Figure 5 Insert tube Figure 6 Locking ring out...
Страница 107: ...107 Figure 14 Install the new water filter Figure 15 Reconnect the water filter Figure 16 Filter installed...
Страница 109: ...109 Figure 17 LCD Display Figure 18 Drain Purge Line Figure 19 Replace Pre filters 1 MICRON 0 45 MICRON...
Страница 111: ...111 Figure 20 Printer compartment Figure 21 Advance paper Figure 22 Feed paper through slot Figure 23 Install new roll...
Страница 115: ...115 Figure 28 Filter locations Figure 29 Disconnect the filter Figure 30 Verify inlet side...
Страница 119: ...119 Figure 33 Fuse locations...
Страница 121: ...121 Figure 34 Connect pressure gauge Figure 35 Bleed some air...
Страница 140: ...140 DSD EDGE W O RECIRC W ADDITIONAL ISOLATION VALVE W CHAMBER FLOW SWITCH HYDRAULIC SCHEMATIC REVISION 02...
Страница 141: ...141...
Страница 142: ...142...
Страница 143: ...143...
Страница 149: ...149...
Страница 150: ...150...
Страница 151: ...151...