6
Safety
This section outlines general safety guidelines for proper operation and service of the reprocessor. Failure to follow these
guidelines may result in severe injury or death to the patient and/or operator. Read and understand all operating and service
procedures before attempting to operate the reprocessor. If the equipment is not used as specified, the protection by the
equipment may be impaired.
Intended Use
Only properly trained individuals may operate or service the reprocessor. Never use the reprocessor for any purpose
other than the manufacturer’s specific intended purpose. It is the responsibility of the facility to maintain and ensure that
adequate training is provided to operators. It is recommended that the facility conduct regular training of all personnel
concerned with the operation and maintenance of this equipment, including emergency procedures for toxic, flammable or
pathogenic material released into the environment. Attendance records of the training should be maintained and evidence
of understanding demonstrated.
Operator Safety
Avoid biological contamination and chemical burns–always wear appropriate personal protective equipment when
handling endoscopes or disinfectant solutions. Never open the reprocessor lid or remove the floating basin lid during
operation.
Moving the DSD EDGE
®
Endoscope Reprocessor
Before moving the DSD EDGE Endoscope Reprocessor, ensure the electrical cord, drain line and water supply line are
either disconnected or are appropriate lengths to accommodate the relocation of the machine. Failure to do so may result
in damage to the machine. While moving the DSD EDGE Endoscope Reprocessor, ensure the machine remains in an
upright position. Moving or resting the machine in any orientation other than an upright position may result in damage to
the machine. For short distances, the DSD EDGE Endoscope Reprocessor may be slid upon it base. Take precautions to
ensure the machine does not tip over which could result in damage to the machine or personal injury. For longer distances,
the machine may be placed upon a dolly or pallet. When loading or unloading the machine onto or off of a dolly or pallet,
utilize appropriate lifting equipment or manpower to avoid damage to the machine or personal injury.
Note: To avoid injury or death from an electrical insulation breakdown within the unit, the
GFI (ground fault interrupter) circuit breaker should be checked for proper operation on an
annual basis.
Note: If during the use of this equipment you see or smell smoke, immediately disconnect
the unit from the power supply, discontinue use and call MEDIVATORS Technical Support at
1-800-444-4729.
Содержание MEDIVATORS DSD EDGE
Страница 1: ...1 Service Manual System Version 5 65...
Страница 23: ...23 Figure 5 Function Keys...
Страница 27: ...27 Figure 6 Set Date Screen...
Страница 29: ...29 Figure 7 Set Time Screens...
Страница 31: ...31 Figure 8 Software Version Screen Flash Code Version or Software version Micro controller Code Version...
Страница 33: ...33 Figure 9 Display Log Screen and sample entry...
Страница 42: ...42 Figure 14 Custom Program Setup Screens P1 Detergent Inject Set Minutes 00 00...
Страница 43: ...43 Figure 15 Custom Program Setup Screen...
Страница 51: ...51 Figure 22 Set Delayed Startup Screen...
Страница 54: ...54...
Страница 96: ...96 Notes...
Страница 101: ...101 Figure 4 Depress locking ring Figure 5 Insert tube Figure 6 Locking ring out...
Страница 107: ...107 Figure 14 Install the new water filter Figure 15 Reconnect the water filter Figure 16 Filter installed...
Страница 109: ...109 Figure 17 LCD Display Figure 18 Drain Purge Line Figure 19 Replace Pre filters 1 MICRON 0 45 MICRON...
Страница 111: ...111 Figure 20 Printer compartment Figure 21 Advance paper Figure 22 Feed paper through slot Figure 23 Install new roll...
Страница 115: ...115 Figure 28 Filter locations Figure 29 Disconnect the filter Figure 30 Verify inlet side...
Страница 119: ...119 Figure 33 Fuse locations...
Страница 121: ...121 Figure 34 Connect pressure gauge Figure 35 Bleed some air...
Страница 140: ...140 DSD EDGE W O RECIRC W ADDITIONAL ISOLATION VALVE W CHAMBER FLOW SWITCH HYDRAULIC SCHEMATIC REVISION 02...
Страница 141: ...141...
Страница 142: ...142...
Страница 143: ...143...
Страница 149: ...149...
Страница 150: ...150...
Страница 151: ...151...