Air pressure switch harness disconnected...........
Check the harness for proper connection. Also, check for
damage, loose wires.
Water line valve not opening ..............................
Repair or replace the component.
Chamber valve not opening ................................
Repair or replace the component.
Air manifold not operating .................................
Repair or replace the component.
Valve drive board defective ................................
Check the valve drive board. Replace if necessary.
Harness disconnected..........................................
Check the harness connections. Replace defective harness.
Air trapped in filter housing................................
Bleed air from the housing.
Slow disinfectant fill into basin.
Air pressure insufficient to open valves..............
Verify the compressor is working correctly. Check the
external air regulator on non-compressor reprocessors.
Contact your Facility Maintenance representative.
Disinfectant screen plugged ...............................
Clean the filter.
Air pressure switch harness disconnected...........
Check the harness for proper connection. Also,
check for damage, loose wires.
Disinfectant supply valve not opening................
Repair or replace the component.
Chamber valve not opening ................................
Repair or replace the component.
Air manifold not operating .................................
Repair or replace the component.
Valve drive board defective ................................
Check the valve drive board. Replace if necessary.
Harness disconnected..........................................
Check the harness connections. Replace defective harness.
Water flows constantly from hookups.
Water valve not sealing correctly........................
Shut off the water supply and contact customer support.
Water does not drain from basin during flush or rinse.
Drain valve is defective.........................................
Repair or replace the drain valve.
External drain line kinked...................................
Repair the drain line.
Drain line plugged ..............................................
Clean the drain line or replace the line, if necessary.
Basin drain screen plugged .................................
Clean the basin screen drain.
Scopes still wet at end of cycle.
Insufficient air flow.............................................
Check the compressor for proper operation.
Air valve not opening .........................................
Repair or replace the component.
Compressor vent valve open/leaking..................
Repair or replace the vent valve.
Chamber valves leaking......................................
Repair or replace the component.
Reprocessor lid does not open on units with optional lidlocks.
Lidlock remains activated...................................
Press the “STOP” button and enter the service code to deactivate
the lidlocks.
Lidlock defective ................................................
Replace the lidlock.
Harness disconnected/defective..........................
Check the connection. Replace a defective harness.
Digital I/O board defective .................................
Replace the I/O board.
24/28V power supply defective ..........................
Replace the power supply.
Содержание MEDIVATORS DSD EDGE
Страница 1: ...1 Service Manual System Version 5 65...
Страница 23: ...23 Figure 5 Function Keys...
Страница 27: ...27 Figure 6 Set Date Screen...
Страница 29: ...29 Figure 7 Set Time Screens...
Страница 31: ...31 Figure 8 Software Version Screen Flash Code Version or Software version Micro controller Code Version...
Страница 33: ...33 Figure 9 Display Log Screen and sample entry...
Страница 42: ...42 Figure 14 Custom Program Setup Screens P1 Detergent Inject Set Minutes 00 00...
Страница 43: ...43 Figure 15 Custom Program Setup Screen...
Страница 51: ...51 Figure 22 Set Delayed Startup Screen...
Страница 54: ...54...
Страница 96: ...96 Notes...
Страница 101: ...101 Figure 4 Depress locking ring Figure 5 Insert tube Figure 6 Locking ring out...
Страница 107: ...107 Figure 14 Install the new water filter Figure 15 Reconnect the water filter Figure 16 Filter installed...
Страница 109: ...109 Figure 17 LCD Display Figure 18 Drain Purge Line Figure 19 Replace Pre filters 1 MICRON 0 45 MICRON...
Страница 111: ...111 Figure 20 Printer compartment Figure 21 Advance paper Figure 22 Feed paper through slot Figure 23 Install new roll...
Страница 115: ...115 Figure 28 Filter locations Figure 29 Disconnect the filter Figure 30 Verify inlet side...
Страница 119: ...119 Figure 33 Fuse locations...
Страница 121: ...121 Figure 34 Connect pressure gauge Figure 35 Bleed some air...
Страница 140: ...140 DSD EDGE W O RECIRC W ADDITIONAL ISOLATION VALVE W CHAMBER FLOW SWITCH HYDRAULIC SCHEMATIC REVISION 02...
Страница 141: ...141...
Страница 142: ...142...
Страница 143: ...143...
Страница 149: ...149...
Страница 150: ...150...
Страница 151: ...151...