120
LEAK TESTER OPTION–CHECK
1. Connect pressure gauge to leak test outlet in “A” upper chamber.
2. Test the Leak Test Board.
• Select side “A”, SETUP 88,code 135
• 51 ENTER, 54 ENTER to setup test
• 95 ENTER. Display will show number 3C.
• Gradually increase the pressure using the inflator. As the pressure increases, the display will read:
- 34 if pressure is greater than 1 psi -31 if pressure is greater than 3 psi
• Gradually decrease the pressure using the toggle on the back of the inflator. As the pressure
decreases, the display will read: -35 if pressure is less than 3 psi -3F if pressure is less than 1 psi
• Press CANCEL.
3. Test Leak Tester system.
• Press 56 ENTER to verify 3 psi control pressure.
• Press 57 ENTER to verify 1 psi hold pressure.
• Bleed some air out of the system using the toggle and ensure that the tester maintains 1 psi.
• Press 50 ENTER to vent the system.
4. Internal Leak Test.
• Press 56 ENTER.
• Allow the pressure to stabilize.
• Press 57 ENTER.
• Monitor pressure for a minimum of 5 minutes. The pressure must not increase or decrease more than 4
mmHg.
• Record the pressure and time on the test report.
• Press 50 ENTER to vent the system
5. Repeat for Station “B”.
• During the test of the Leak Test Board for Station “B”, the display will show the
following readings:
- As the pressure increases, the display will read: 1C if pressure is greater than 1 psi, 8C if
pressure is greater than 3 psi.
- As the pressure decreases, the display will read: 9C if pressure is less than 3 psi, FC if pressure is less
than 1 psi.
Содержание MEDIVATORS DSD EDGE
Страница 1: ...1 Service Manual System Version 5 65...
Страница 23: ...23 Figure 5 Function Keys...
Страница 27: ...27 Figure 6 Set Date Screen...
Страница 29: ...29 Figure 7 Set Time Screens...
Страница 31: ...31 Figure 8 Software Version Screen Flash Code Version or Software version Micro controller Code Version...
Страница 33: ...33 Figure 9 Display Log Screen and sample entry...
Страница 42: ...42 Figure 14 Custom Program Setup Screens P1 Detergent Inject Set Minutes 00 00...
Страница 43: ...43 Figure 15 Custom Program Setup Screen...
Страница 51: ...51 Figure 22 Set Delayed Startup Screen...
Страница 54: ...54...
Страница 96: ...96 Notes...
Страница 101: ...101 Figure 4 Depress locking ring Figure 5 Insert tube Figure 6 Locking ring out...
Страница 107: ...107 Figure 14 Install the new water filter Figure 15 Reconnect the water filter Figure 16 Filter installed...
Страница 109: ...109 Figure 17 LCD Display Figure 18 Drain Purge Line Figure 19 Replace Pre filters 1 MICRON 0 45 MICRON...
Страница 111: ...111 Figure 20 Printer compartment Figure 21 Advance paper Figure 22 Feed paper through slot Figure 23 Install new roll...
Страница 115: ...115 Figure 28 Filter locations Figure 29 Disconnect the filter Figure 30 Verify inlet side...
Страница 119: ...119 Figure 33 Fuse locations...
Страница 121: ...121 Figure 34 Connect pressure gauge Figure 35 Bleed some air...
Страница 140: ...140 DSD EDGE W O RECIRC W ADDITIONAL ISOLATION VALVE W CHAMBER FLOW SWITCH HYDRAULIC SCHEMATIC REVISION 02...
Страница 141: ...141...
Страница 142: ...142...
Страница 143: ...143...
Страница 149: ...149...
Страница 150: ...150...
Страница 151: ...151...