137
Phase
Option
Description
Default
Typical
Min.
Limit
Max.
Limit
Setting
State
#
Function
#
Rinse 2
Endoscope channel flush with water
30 sec
30 sec
0 sec
999 sec
Factory Set
51
Diag. 63
Basin fill minimum (level sensor not monitored)
90 sec
90 sec*
0 sec
999 sec
Factory Set
100
Diag. 62
Basin fill balance (sensor monitored)
5 min
10 sec*
Hard Coded
101
Top off
30 sec
30 sec
0 sec
999 sec
Factory Set
102
Diag. 67
Replenish with fresh water
0 sec
0 sec
0 sec
99m:59s
User Set
103
Setup 5
Rinse 2
Drain the basin while flushing the scope
channels
30 sec
30 sec
0 sec
999 sec
Factory Set
96
Diag. 61
Drain
30 sec
30 sec
Hard Coded
97
Air purge through chamber line
5 sec
5 sec
Hard Coded
92
Endoscope channel air purge
30 sec
30 sec
0 sec
999 sec
Factory Set
94
Diag. 63
Clear air lines
5 sec
5 sec
Hard Coded
95
Rinse 3
Endoscope channel flush with water
30 sec
30 sec
0 sec
999 sec
Factory Set
53
Diag. 63
Basin fill minimum (level sensor not monitored)
90 sec
90 sec*
0 sec
999 sec
Factory Set
100
Diag. 62
Basin fill balance (sensor monitored)
5 min
10 sec*
Hard Coded
101
Top off
30 sec
30 sec
0 sec
999 sec
Factory Set
102
Diag. 67
Replenish with fresh water
0 sec
0 sec
0 sec
99m:59s
User Set
103
Setup 5
Rinse 2
Drain the basin while flushing the scope
channels
30 sec
30 sec
0 sec
999 sec
Factory Set
96
Diag. 61
Drain
30 sec
30 sec
Hard Coded
97
Air purge through chamber line
5 sec
5 sec
Hard Coded
92
Endoscope channel air purge
30 sec
30 sec
0 sec
999 sec
Factory Set
94
Diag. 63
Clear air lines
5 sec
5 sec
Hard Coded
95
Alcohol
Alcohol dispensing into manifold
10 sec
10 sec
0 sec
59 sec
User Set
55
Setup 5
Alcohol Inject
Endoscope channel alcohol purge
0 sec
0 sec
0 sec
99m:59s
User Set
56
Setup 5
Alcohol
Clear air lines
5 sec
5 sec
Hard Coded
57
Air
Air purge through chamber lines
5 sec
5 sec
Hard Coded
58
Endoscope channel air purge
3 min
3 min
0 sec
99m:59s
User Set
60
Setup 5
Air
Clear air lines
5 sec
5 sec
Hard Coded
61
1 - The time in the Typical column where an * symbol is indicated, is subject to how fast the basin fills. An alarm will occur when default time is reached.
2 - Selecting zero time during Setup 5 will cancel all accompanying states in the phase.
3 - The “FLUSH” message is displayed in the Phase column indicates that the state is only active when the SOAK time is set to zero.
4 - States with a leak test option are active if the unit is provided with the Leak Tester and the sheath sensor is enabled in the diagnostics mode.
Содержание MEDIVATORS DSD EDGE
Страница 1: ...1 Service Manual System Version 5 65...
Страница 23: ...23 Figure 5 Function Keys...
Страница 27: ...27 Figure 6 Set Date Screen...
Страница 29: ...29 Figure 7 Set Time Screens...
Страница 31: ...31 Figure 8 Software Version Screen Flash Code Version or Software version Micro controller Code Version...
Страница 33: ...33 Figure 9 Display Log Screen and sample entry...
Страница 42: ...42 Figure 14 Custom Program Setup Screens P1 Detergent Inject Set Minutes 00 00...
Страница 43: ...43 Figure 15 Custom Program Setup Screen...
Страница 51: ...51 Figure 22 Set Delayed Startup Screen...
Страница 54: ...54...
Страница 96: ...96 Notes...
Страница 101: ...101 Figure 4 Depress locking ring Figure 5 Insert tube Figure 6 Locking ring out...
Страница 107: ...107 Figure 14 Install the new water filter Figure 15 Reconnect the water filter Figure 16 Filter installed...
Страница 109: ...109 Figure 17 LCD Display Figure 18 Drain Purge Line Figure 19 Replace Pre filters 1 MICRON 0 45 MICRON...
Страница 111: ...111 Figure 20 Printer compartment Figure 21 Advance paper Figure 22 Feed paper through slot Figure 23 Install new roll...
Страница 115: ...115 Figure 28 Filter locations Figure 29 Disconnect the filter Figure 30 Verify inlet side...
Страница 119: ...119 Figure 33 Fuse locations...
Страница 121: ...121 Figure 34 Connect pressure gauge Figure 35 Bleed some air...
Страница 140: ...140 DSD EDGE W O RECIRC W ADDITIONAL ISOLATION VALVE W CHAMBER FLOW SWITCH HYDRAULIC SCHEMATIC REVISION 02...
Страница 141: ...141...
Страница 142: ...142...
Страница 143: ...143...
Страница 149: ...149...
Страница 150: ...150...
Страница 151: ...151...