130
Reset Alarms? ........................ Prompts the user to press the enter button to cancel an alarm
Sheath Fail ............................. Indicates that the sheath test failed
Shth Sen Err ........................... Indicates that the sheet tester measured pressure at the start of a cycle
SNVRAM Err ............................ Short Non-Volatile RAM error, can only be cleared in Diagnostics 70
SSG A low ............................... Alarm indicating that the SSG has timed out before filling part A reservoir (Part A
...............................................
chemistry low)
SSG B low................................Alarm indicating that the SSG has timed out before filling part B reservoir (Part
................................................
B chemistry low)
SSG A Sen Er ...........................Alarm indicating that the SSG dosing reservoir A sensors are reading incorrectly
................................................
(sensor hardware error)
SSG B Sen Er ...........................Alarm indicating that the SSG dosing reservoir B sensors are reading
................................................
incorrectly (sensor hardware error)
SSG Cmd Err ............................Alarm indicating that the SSG has received an unknown command from the
................................................
DSD (software error?)
SSG Comm Err .........................Alarm indicating that the SSG detected a failure in the DSD Communications
SSG CRC Fail
Alarm indicating that the SSG failed its internal Software Checksum test
(program not functioning correctly)
SSG Data Er ............................ Alarm indicating that the SSG Database was out of range. DSD and SSG data
...............................................
do not match.
SSG Dose Err .......................... Alarm indicating that the SSG is not indicating completion prior to DSD step
...............................................
advancing (SSG lost database or program step error)
SSG Fill Err ............................. Alarm indicating that the DSD has advanced to the next step prior to SSG fill
...............................................
completion ( program step error)
SSG HW Err ............................ Alarm indicating that the SSG has experienced a hardware error (includes
...............................................
capacitive sensor missing)
SSG Missing ........................... Alarm indicating the DSD detected a failure in the SSG Communication (DSD
...............................................
doesn’t see the SSG)
SSG Rsvr Err ........................... Alarm indicating that an SSG dosing reservoir sensor has remained high (not
...............................................
empty) after disinfect cycle
Sta Not Idle ............................. Indicates that the selected operation can not be performed because the station
...............................................
is not currently idle
Sth Disabled ........................... Indicates that the sheath (leak) tester was disabled
Task Unfin ............................... Alarm indicating that the SSG is reporting a Task Not Finished Error ; fill or dose
...............................................
did not complete
Time Err .................................. Time base error, can only be cleared in Diagnostics 70
Содержание MEDIVATORS DSD EDGE
Страница 1: ...1 Service Manual System Version 5 65...
Страница 23: ...23 Figure 5 Function Keys...
Страница 27: ...27 Figure 6 Set Date Screen...
Страница 29: ...29 Figure 7 Set Time Screens...
Страница 31: ...31 Figure 8 Software Version Screen Flash Code Version or Software version Micro controller Code Version...
Страница 33: ...33 Figure 9 Display Log Screen and sample entry...
Страница 42: ...42 Figure 14 Custom Program Setup Screens P1 Detergent Inject Set Minutes 00 00...
Страница 43: ...43 Figure 15 Custom Program Setup Screen...
Страница 51: ...51 Figure 22 Set Delayed Startup Screen...
Страница 54: ...54...
Страница 96: ...96 Notes...
Страница 101: ...101 Figure 4 Depress locking ring Figure 5 Insert tube Figure 6 Locking ring out...
Страница 107: ...107 Figure 14 Install the new water filter Figure 15 Reconnect the water filter Figure 16 Filter installed...
Страница 109: ...109 Figure 17 LCD Display Figure 18 Drain Purge Line Figure 19 Replace Pre filters 1 MICRON 0 45 MICRON...
Страница 111: ...111 Figure 20 Printer compartment Figure 21 Advance paper Figure 22 Feed paper through slot Figure 23 Install new roll...
Страница 115: ...115 Figure 28 Filter locations Figure 29 Disconnect the filter Figure 30 Verify inlet side...
Страница 119: ...119 Figure 33 Fuse locations...
Страница 121: ...121 Figure 34 Connect pressure gauge Figure 35 Bleed some air...
Страница 140: ...140 DSD EDGE W O RECIRC W ADDITIONAL ISOLATION VALVE W CHAMBER FLOW SWITCH HYDRAULIC SCHEMATIC REVISION 02...
Страница 141: ...141...
Страница 142: ...142...
Страница 143: ...143...
Страница 149: ...149...
Страница 150: ...150...
Страница 151: ...151...