10.3.4 VIBROCONTROL 1870 Terminals
Non-contacting vibration displacement sensors
The 3-wire displacement sensors are connected to the following terminals:
Signal (Displacement 1-5) and GND (shared) and voltage (-24 V shared). Up to 5 displacement
sensors can be connected to a VC-1870.
Vibration displacement sensors
13 (Displacement 1) / 15 (-24 V) / 14 (GND)
16 (Displacement 2) / 15 (-24 V) / 14 (GND)
17 (Displacement 3) / 19 (-24 V) / 18 (GND)
20 (Displacement 4) / 19 (-24 V) / 18 (GND)
21 (Displacement 5) / 19 (-24 V) / 22 (GND)
The sensor inputs "Displacement 1-4" for monitoring the relative
shaft vibration.
The sensor input "Displacement 5" is used exclusively for the axial
position!
Relative shaft vibration:
Axial position: