NOTE!
The VC 18xx devices are equipped with resistors or
diodes at the terminals for sensors. If the inputs are not
used, the resistors must remain connected. In addition,
disable any sensor channels not being used with the
Compact Setup software.
Before connecting sensors, remove the resistors or
diodes from the relevant terminals.