Page 26 of 110
© Brüel & Kjær Vibro
● C106834.002 / V06
Technical alterations reserved!
EN
VIBROCONTROL 1870 with non-contacting displacement sensors:
Enter the sensitivity of the connected displacement sensor. It is indicated by the sensor
manufacturer. A typical value for standard displacement sensor is 8 V/mm.
The adjustment of the displacement sensor also sets up the distance measurement (GAP) (if
enabled). This also makes it possible to check whether the sensor is working in its defined
measuring range.
The sensitivity of the sensor is normally given in V/mm or mV/µm.
Low-level and High-level alarms are available for distance measurements (GAP alarms) to indicate
that something is wrong with the setup of the displacement sensor.