54
5.
Select
Usmin
…
and press enter. The user can choose a Usmin value between
1 V
–
80 V
.
Once a value is selected, press
to confirm.
6.
Use the
key to select
Off/On
and use
keys to toggle between the two options.
Choose
On
to be able to run the function.
7.
Once both parameters are set, press
to confirm the changed settings and press
several times to exit the menu. If the Starting profile function was setup properly,
the
Trig
indicator will be highlighted and
Profile for the reset test
will be shown on the
front panel, shown below.
8.
To enable the Profile reset test function, press the
key, on the front panel. The
proper voltage selected should be metered on the front panel display.
9.
To run the Profile reset test function press
and
to trigger the test to run.
Please be aware this test does run for 340 seconds (See profile reset test figure above).
Starting profile
There are different profiles available for 12 V systems and 24 V systems, please refer to the
tables above to configure the power supply options properly.
Follow the steps below to set up the ISO16750-2:2012(E) for Starting profile function:
1.
Access the main menu by pressing
and the
.
2.
Use the key to select
FUNC.
and press
.
3.
Use the key to select
ISO16750-2
and press
. The screen below will display:
4.
While
Short voltage drop
is flashing, use keys to toggle to
Starting profile
(shown
below)
and press
to access that option.
ISO16750-2 profile
P rof il e fo r th e reset test
0.000V 0.000A
0.0 W
P rof il e for reset
Trig