43
9.
This is prompting the user to enter the voltage setting for
step 1
. Use the voltage and
current adjust knob or the numeric keypad and enter
1.000
. Then press
.
10.
The display will change to the following, prompting the user to enter the current setting
for
step 1
.
11.
Similar to entering the voltage setting in the previous step, do the same to enter
1.000
as the current setting. Then press
. Now, the following will display:
12.
Enter the width, which is the time to hold the voltage and current setting configured in
the previous steps. Enter
1.000
(for 1 second). Press
. Now, the following will
display:
13.
Enter the slope, which defines the voltage slew rate (transition time between steps) for
step 1
. Enter
0.005
. Press
.
14.
Now, the display will return back to the display shown in step number 8, except it will
prompt the user to enter the voltage setting for
step 2
of the sequence. Follow steps 8-
13 to configure step 2 with: 2V, 1A, 2s width, 0.005s slope. Then step 3 with: 3V, 1A, 3s
width, 0.005s slope. Then step 4 with: 1.5V, 1A, 3s width, 0.005s slope. Finally, step 5
with: 2.5V, 1A, 4s width, 0.005s.
15.
After
step 5
parameters are configured, the display will show the following to request
confirmation to save the step parameters into memory. Press key to select
Yes
and
press
.
EDIT SEQ CURRENT
S eq Step 1 Cu rrent = 1.000 A
EDIT SEQ WIDTH
S eq Step 1 Width = 1. 000 s
EDIT SEQ SLOPE
S eq Step 1 Slope = 0.005 s