23
Power-On
Rst
Load Setup
Off
External Control
10v-M
Limit
Vmin = 0.000V, Vmax = 81.000V
Online Setup
Off
Output
Disabled
Configure Power-On State
The initial power-On state of the power supply can be configured by following the steps below:
1.
From the
SYSTEM
menu, select
Power-On
and press
.
2.
There are two options:
Rst(Def)
–
Factory Default.
Sav0
–
Settings before last power up.
3.
Select the settings you want during power up, and press
to save changes.
4.
To exit the menu at any time, press
twice.
Configure Trigger Source
The trigger function is used to initiate the start of running a program sequence (list). The trigger
source can be set so that users can send a trigger from the front panel or through a remote
command via remote interface. Follow the below steps to configure the trigger mode:
1.
From the
SYSTEM
menu, browse and select
Trigger
and press
.
2.
There are two options:
Manual(Def)
–
Manual trigger. Front panel trigger button is used to send a trigger
(press
and
to send trigger).
Bus
–
Bus trigger. Remote command
*TRG
is used to send a trigger.
3.
To exit the menu at any time, press
twice.