Model:  9115/B, 9115/B-AT, 9116/B, 9117 

Multi-Range DC Power Supply

 

USER MANUAL 

Содержание 9115/B

Страница 1: ...Model 9115 B 9115 B AT 9116 B 9117 Multi Range DC Power Supply USER MANUAL...

Страница 2: ...ctrical energy available and the voltage impulses that may occur on electrical conductors associated with these category ratings The category rating is a Roman numeral of I II III or IV This rating is...

Страница 3: ...imize shock hazard the instrument chassis and cabinet must be connected to an electrical safety ground This instrument is grounded through the ground conductor of the supplied three conductor AC line...

Страница 4: ...res are within specifications In direct sunlight This instrument is intended to be used in an indoor pollution degree 2 environment The operating temperature range is 0 C to 40 C and the operating hum...

Страница 5: ...th the power cord removed dangerous voltages may exist when the covers are removed To avoid injuries always disconnect the power cord from the instrument disconnect all other connections for example t...

Страница 6: ...This instrument contains one or more cooling fans For continued safe operation of the instrument the air inlet and exhaust openings for these fans must not be blocked nor must accumulated dust or othe...

Страница 7: ...to the instrument Do not place a hot soldering iron on the instrument Do not pull the instrument with the power cord connected probe or connected test lead Do not move the instrument when a probe is...

Страница 8: ...C of the European Parliament and the Council of the European Union on waste electrical and electronic equipment WEEE and in jurisdictions adopting that Directive is marked as being put on the market a...

Страница 9: ...ents of 2006 95 EC Low Voltage Directive and 2004 108 EC Electromagnetic Compatibility Directive with the following standards Low Voltage Directive EN61010 1 2001 EMC Directive EN 61000 3 2 2006 EN 61...

Страница 10: ...ins connect disconnect switch on the front of the instrument DC current Alternating current Fuse Symbol Chassis earth ground symbol Ground terminal Protective earth ground CAUTION indicates a hazardou...

Страница 11: ...escription 6 1 6 Display Overview 7 Display Description 7 2 Getting Started 8 2 1 Input Power and Fuse Requirements 8 Input Power 8 Fuse Requirements 11 Fuse Replacement 11 2 2 Output Connections 12 2...

Страница 12: ...Menu 29 Load Setup Option 29 External Analog Control 30 Configure Voltage Limit 35 Parallel Series Connection 36 Connection and Setup 36 3 5 Program Sequence Mode List mode 39 Configure Sequence Para...

Страница 13: ...xii 4 2 Remote Commands 60 5 Troubleshooting Guide 61 General 61 Remote Control 61 6 Specifications 62 7 Calibration 63 SERVICE INFORMATION 64 LIMITED ONE YEAR WARRANTY 65...

Страница 14: ...mA 2mV 3mA for model 9117 Standard USBTMC GPIB non B series RS232 RS485 and Ethernet model 9117 only External analog interface with control and monitoring functions Master Slave Mode for series and pa...

Страница 15: ...inch rackmount and is of 1U size The 9117 power supply s dimensions WxHxD are approximately 436 5 mm x 89 3 mm x 532 mm 17 19 x 3 52 x 20 95 It is also designed to fit in a standard 19 inch rackmount...

Страница 16: ...3 Figure 1 3 Front View for 9117 Figure 1 4 Side View for 9117...

Страница 17: ...and selecting menu items On Off Lock button Control the output state or locks the front panel button Save Recall button Used to save and recall instrument settings Meter display toggle Local button U...

Страница 18: ...et P max button Used to configure output current or set the maximum power output limit OVP set button Used to configure over voltage protection OVP value V set Slope button Used to configure output vo...

Страница 19: ...minal Rear Cooling Fan GPIB non B MODELS ONLY Interface USB Interface RS 232 Interface AC Input Receptacle Fuse Box RS 485 Interface Remote Sense Terminals Ethernet LAN Interface Model 9117 only 1 3 1...

Страница 20: ...onstant voltage CV operation Rmt Indicates remote mode Addr Indicates remote communication activity Error Indicates an error has occurred Trig Indicates waiting for trigger for list operation Prot Ind...

Страница 21: ...safety certified for this instrument operating in rated range To change a cable or add an extension cable be sure that it can meet the required power ratings for this instrument Any misuse with wrong...

Страница 22: ...1 Remove the safety metal housing held in place with one screw to reveal the 3 wire AC connector port on the rear panel of the power supply 2 Connect the input receptacle green terminal block to the i...

Страница 23: ...10 Figure 2 1 AC Power cable housing assembly Follow the instructions below to connect the AC power cord to the AC supply mains distribution Step 1 Steps 2 3 Step 6 Step 4 5...

Страница 24: ...s Model Fuse Description 9115 B 15AT 250V 6 x 32 mm slow blow ceramic tube fuse 9115 B AT 15AT 250V 6 x 32 mm slow blow ceramic tube fuse 9116 B 15AT 250V 6 x 32 mm slow blow ceramic tube fuse 9117 20...

Страница 25: ...y 4 size screwdriver Note The screws on the terminals can be completely removed to allow for ring type adapters must be greater than 6mm in diameter Before connecting wires to the output terminals tur...

Страница 26: ...ord to the AC receptacle in the rear panel and press the power switch to the ON position to turn ON the instrument It will run through a self test procedure with the screen shown in below Self test Er...

Страница 27: ...eck basic current output of the power supply 1 Turn on the power supply The display will show the OFF annunciator above the voltage display Be sure that the output is disabled the should not be lit wh...

Страница 28: ...e command It can also be found from within the menu system by following these steps 1 Press and press to enter the menu system 2 Press the arrow three times until INFO is blinking Press and the displa...

Страница 29: ...d Enables dummy load to increase speed of voltage fall time Ext Ctrl Configures external analog control Limit Configures voltage setting limits Online Configures series parallel connection and master...

Страница 30: ...to access the selected menu item 7 There may be parameters or options to select within each menu item Follow the same instructions as described in the previous steps to select them To save changes to...

Страница 31: ...e and press to set the current 3 To change the cursor position to adjust with the voltage adjust knob use the keys to move left or right Note To see the changes to the setting current be sure the disp...

Страница 32: ...k hazard under high voltage output conditions Remote Sense To enable remote sense follow the steps below 1 Power OFF the supply and disconnect all loads and cables connected to it 2 Use a small flat b...

Страница 33: ...bottom right Use the voltage adjust knob or the numeric keypad to enter the OVP limit 2 Press to save the change The display will then show TOVPdly This parameter specifies the delay time to which to...

Страница 34: ...t adjust knob or the numeric keypad to enter the maximum power output limit 2 Press to save the change 3 Use the voltage or current adjust knob or the numeric keypad to enter a value This can be from...

Страница 35: ...g the instrument to factory default will change all current instrument settings and parameters back to their default values 1 From the SYSTEM menu select Initialize and press 2 The instrument will ret...

Страница 36: ...nges 4 To exit the menu at any time press twice Configure Trigger Source The trigger function is used to initiate the start of running a program sequence list The trigger source can be set so that use...

Страница 37: ...ing screen will appear 2 Use the current adjust knob or the numeric keypad to enter the storage group Select between 0 9 Press to save selection 3 To exit the menu at any time press twice Save Setting...

Страница 38: ...e cursor on the display will reappear indicating that the instrument is no longer in Recall mode Enable Disable Key Sound The instrument initially has key sound enabled from factory To disable or re e...

Страница 39: ...sts the options that can be changed for each setting Baudrate 4800 9600 19200 38400 57600 115200 Data bits 8 Parity N None E Even O Odd Stop bit 1 2 Addr 0 to 31 Note The default is 4800 8 N 1 Address...

Страница 40: ...Use the current adjust knob or the numeric keypad to enter an address from 0 31 5 Press to save the selected address and the display will return to the Communication menu 6 Press several times to exi...

Страница 41: ...upply that you want to connect together and control making sure that each of them have a different address assigned 9 Press several times to exit the menu Ethernet Model 9117 only The Ethernet LAN int...

Страница 42: ...eter and press 2 Select between the two options Off Def Return Meter function disabled Delay Enables return meter function 3 Select the desired option and press to save the change 4 To exit the menu a...

Страница 43: ...ummy load To enable or disable this function follow the steps below 1 From the CONFIG menu select Load and press 2 Select between the two options Off Disables the dummy load default On Enables the dum...

Страница 44: ...current programming is on Pin 22 of the DB25 Analog interface 1 From the CONFIG menu browse and select Ext Ctrl and press The following screen will display 2 Select the second item 10v 10k P or 5v 5k...

Страница 45: ...output R P Use external resistance for programming the output 3 Press when finished making the selection 4 Press several times to exit the menu Enable Disable External Analog Control If external anal...

Страница 46: ...ing input 21 REF_10V 10V DC reference output 22 IPRG Current programming input 23 VMON Voltage monitoring output 24 IMON Current monitoring output 13 AGND Ground pin DO NOT input any voltages above 10...

Страница 47: ...ly status can be monitored using POWER_OK pin 16 output and DGND pin 1 During normal operating conditions this pin will output 5 VDC If OTP condition or other abnormalities occur this pin will output...

Страница 48: ...e Limit The minimum and maximum set voltage can be configured on the power supply to limit the settable voltage range for protection Note At any time during operation when you are unable to set to a d...

Страница 49: ...o function Table 3 2 Series Parallel Configuration Table Model Series Parallel 9115 B and 9115 B AT 3 units 240 V Max 3 units 180 A Max 9116 B 2 units 300 V Max 4 units 120 A Max 9117 3 units 240 V Ma...

Страница 50: ...s positive terminal to the negative terminal of another Do the same for all the power supplies Then connect all of the power supplies Pin 1 of the RS 485 interface together Do the same for Pin 5 Below...

Страница 51: ...ave Configuration After physically connecting the power supplies for parallel or series operation power ON the power supplies and follow the steps below to configure a power supply as a master or slav...

Страница 52: ...is not available on model 9117 List mode functionality is only available in the software program for model 9117 The power supply models 9115 B AT and 9116 B only has a list mode feature that allows s...

Страница 53: ...ations Additionally the output of the power supply must be disabled prior to setting up the configurations Configure Sequence Parameters Each sequence can store from 1 to 10 steps Each step s paramete...

Страница 54: ...a sequence with fixed current limit of 1 A as illustrated below 1 Access the main menu by pressing and the 2 Use the key to select LIST FUNC then press twice for model 9115 B AT and press The below s...

Страница 55: ...t to configure as active use the numeric keypad to select so that the corresponding number changes to Y Pressing the same number will change the Y back to the corresponding number For example if you w...

Страница 56: ...isplay 13 Enter the slope which defines the voltage slew rate transition time between steps for step 1 Enter 0 005 Press 14 Now the display will return back to the display shown in step number 8 excep...

Страница 57: ...uence 1 and sequence 2 to run after sequence 1 Suppose sequence 2 is already stored into memory and is illustrated as the following This example setup will run sequence 1 two times and sequence 2 thre...

Страница 58: ...mes is 65535 times 5 The next display will show the following 6 This is similar to step 5 and 6 in the previous section for Configure Sequence Parameters Each of the numbers on the display represents...

Страница 59: ...10 Now it will display the following to confirm saving the program list 11 Press key to select Yes and press to save the program list into memory 12 If stored successfully the display will show a mess...

Страница 60: ...indicator will appear at the top middle of the display 2 Now press twice to exit the menu The following will display 3 If the display is showing the measured voltage and current value the backlight sh...

Страница 61: ...e the ability to test 12 volt and 24 volt systems For details on the DIN 40839 and ISO 16750 2 2012 E standards please see www din de DIN standards and www iso org ISO standards DIN 40839 Parameters T...

Страница 62: ...ys to toggle between the two options Set to 12V for the 12V test function and 24V for the 24V test function 5 Use the key to select Off On and use keys to toggle between the two options Choose On to b...

Страница 63: ...o run as many times as needed ISO 16750 2 2012 E Parameters The ISO16750 2 2012 E function contains three different test simulations Short voltage drop Profile for the reset test and Starting profile...

Страница 64: ...51 Figure 3 10 Profile for the reset test waveform Figure 3 11 Starting profile waveform...

Страница 65: ...ues for 24 V system devices Short voltage drop Follow the steps below to set up the ISO16750 2 2012 E for Short voltage drop function 1 Access the main menu by pressing and the 2 Use the key to select...

Страница 66: ...voltage drop function press the key on the front panel The proper voltage 12V or 24V should be metered on the front panel display 8 To run Short voltage drop function press and to trigger the test to...

Страница 67: ...ould be metered on the front panel display 9 To run the Profile reset test function press and to trigger the test to run Please be aware this test does run for 340 seconds See profile reset test figur...

Страница 68: ...10 Once both parameters are set press to confirm the changed settings and press several times to exit the menu If the Starting profile function was setup properly the Trig indicator will be highlight...

Страница 69: ...pose you want to simulate the voltage signal below by manually setting the voltage from 5 V to 10 V over a span of 5 seconds then manually changing it to 1 V and have it decrease from 10 V to 1 V over...

Страница 70: ...e measured voltage and current display 8 Now set voltage to 10 V using the numeric keypad then press Observe that the supply will slowly ramp its voltage output to 10 V over a span of 5 seconds 9 When...

Страница 71: ...ansmit Data 3 Receive Data 4 5 GND 6 7 8 9 A straight pin to pin DB9 female to DB9 male serial cable is required for using the RS 232 interface Do not use a null modem or crossover DB9 serial cable GP...

Страница 72: ...ot used PIN Description 1 A 2 3 4 5 B 6 7 8 9 USBTMC The device is SR1 RL1 and DT1 enabled It can receive the following request REN_CONTROL GO_TO_LOCAL LOCAL_LOCKOUT When it receives MsgID TRIGGER USB...

Страница 73: ...nds These commands enable a computer to remotely communicate and control the power supply over any of the supported remote interfaces USBTMC RS 232 GPIB non B MODELS ONLY RS 485 and Ethernet Model 911...

Страница 74: ...ome my display voltage is lower than the voltage I set This is often because of the resistances from the test leads wires connected between the supply and the DUT device under test which causes a drop...

Страница 75: ...Current 100 mArms 40 mArms 120 mArms Temperature Coefficient 0 C 40 C 2 Voltage 0 02 30 mV Current 0 05 10 mA Readback Temperature Coefficient 0 C 40 C 2 Voltage 0 02 30 mV Current 0 05 5 mA Efficien...

Страница 76: ...on We recommend a calibration interval of once per year to ensure that the power supply meets specifications The instrument features closed case calibration however it is performed remotely via the re...

Страница 77: ...probes connectors and accessories that you are using with the device Customers not on an open account must include payment in the form of a money order or credit card For the most current repair char...

Страница 78: ...strument via our website www bkprecision com Exclusions This warranty does not apply in the event of misuse or abuse of the product or as a result of unauthorized alterations or repairs The warranty i...

Страница 79: ...22820 Savi Ranch Parkway Yorba Linda CA 92887 www bkprecision com 2014 2021 B K Precision Corp Printed in China v020322...