bkSpecto Ultrasound System
Mechanical Parts
SM41406-4
6-25
4. Cut off the zip tie
Caution:
Do not pull at the wires but use the connectors for a firm grip.
5. Remove the four
Speaker wires.
6. Remove the video
connector.
7. Remove the power
connector.
8. Remove the four
M3X8 screws with a
torx #10.
9. Remove the display
controller.
Power
Video
Connector
Speaker
wires