bkSpecto Ultrasound System
Troubleshooting
SM41406-4
4-85
5. Make a note of the
Full computer name
and
Domain
(in this example
b1016.bkmed.dk
and
analogic.com
).
NOTE:
IF the computer is a member of a
Workgroup
rather than a
Domain
then
Domain
will be replaced by
Workgroup.
Configure Microsoft Windows-based PC
NOTE:
This is not to be done on the scanner.
The scanner requires a shared folder on the hard disk where the files can be saved.
To Configure the PC:
1. Start File Explorer.
2. Create a folder for instance
bkSpectoData
at the root of the C:\ drive.
3. Right click on the folder and select
Properties
.
4. Select
Sharing
.
Содержание bkSpecto 1300
Страница 1: ...Service Manual Diagnostic Ultrasound System 1300 bkSpecto...
Страница 2: ......
Страница 4: ......
Страница 18: ...General Information bkSpecto Ultrasound System 1 14 SM41406 4 This page is intentionally left blank...
Страница 94: ...Troubleshooting bkSpecto Ultrasound System 4 72 SM41406 4 6 Finish procedure online or by call Microsoft...
Страница 111: ...bkSpecto Ultrasound System Troubleshooting SM41406 4 4 89...
Страница 115: ...Adjustment Procedure bkSpecto Ultrasound System 5 4 SM41406 4 This page is intentionally left blank...
Страница 142: ...bkSpecto Ultrasound System Mechanical Parts SM41406 4 6 27 3 Remove the two speaker wires 4 Remove 11 M3X8 screws...
Страница 143: ...Mechanical Parts bkSpecto Ultrasound System 6 28 SM41406 4 5 Gently lift and slide out the Speaker Assembly...
Страница 149: ...Mechanical Parts bkSpecto Ultrasound System 6 34 SM41406 4 This page is intentionally left blank...
Страница 165: ...Theory of Operation bkSpecto Ultrasound System 8 12 SM41406 4 This page is intentionally left blank...