bkSpecto Ultrasound System
Troubleshooting
SM41406-4
4-21
Entering Service Mode (with USB keyboard)
8. Turn on the scanner and wait until the boot cycle is complete and opens in
Console
mode
.
9. Press
ALT + X
.
10. Type in the Service Mode Password, chapter 4.28.
11. Press OK.
12. The scanner will restart and boot in
Service mode
.
Note:
In cases where this procedure does not work it is possible to boot from a bootable device. This
is described in section 4.26.
Entering Console Mode
13. On the Service Desktop select Console mode.
14. When the terminal launches type
y
and press
Enter
.
15. The scanner will restart and boot in
Console mode
.
Essential Operation
Several operations are required to complete almost any service task involving system
software.
Opening File Explorer
1. Access
Service mode
, 4.18.1.
2. Press the
Windows
button in the lower left corner.
3. Select File Explorer.
Содержание bkSpecto 1300
Страница 1: ...Service Manual Diagnostic Ultrasound System 1300 bkSpecto...
Страница 2: ......
Страница 4: ......
Страница 18: ...General Information bkSpecto Ultrasound System 1 14 SM41406 4 This page is intentionally left blank...
Страница 94: ...Troubleshooting bkSpecto Ultrasound System 4 72 SM41406 4 6 Finish procedure online or by call Microsoft...
Страница 111: ...bkSpecto Ultrasound System Troubleshooting SM41406 4 4 89...
Страница 115: ...Adjustment Procedure bkSpecto Ultrasound System 5 4 SM41406 4 This page is intentionally left blank...
Страница 142: ...bkSpecto Ultrasound System Mechanical Parts SM41406 4 6 27 3 Remove the two speaker wires 4 Remove 11 M3X8 screws...
Страница 143: ...Mechanical Parts bkSpecto Ultrasound System 6 28 SM41406 4 5 Gently lift and slide out the Speaker Assembly...
Страница 149: ...Mechanical Parts bkSpecto Ultrasound System 6 34 SM41406 4 This page is intentionally left blank...
Страница 165: ...Theory of Operation bkSpecto Ultrasound System 8 12 SM41406 4 This page is intentionally left blank...