Mechanical Parts
bkSpecto Ultrasound System
6-2
SM41406-4
6
Introduction
This section describes how to disassembly the scanner step by step.
bkSpecto is divided into four main mechanical parts:
Engine
Monitor
Top Assembly
Wheelbase
Caution:
Turn off the scanner before replacing mechanical parts and lock the wheels.
6.1
Screw Overview
All screws used in this section are described with part number and size in Table 6-1.
Part
Description
Usage
111056
M3X5
Engine
109428
M3X6
Left chassis
Charger board
Main mounting plate
24-061753
M3X6
Power supply
24-061754
M3X8
Display controller
24-061712
M3X12
IO board
YT0312
M3X12
Probe holder
24-061771
M3X20
Power supply
111055
M3X35
Engine
117657
M4X10
Cover engine
21-05766-01 M4X10
Cover back engine
109430
M4X12
Engine
117268
M4X25
Monitor
110829
M4X30
Top assembly
24-061752
M4X128
Top assembly
109439
M5X12
Top assembly
109434
M5X16
Top assembly
6-83839-01
M5X20
Top assembly
Table 6-1: Screw overview
Содержание bkSpecto 1300
Страница 1: ...Service Manual Diagnostic Ultrasound System 1300 bkSpecto...
Страница 2: ......
Страница 4: ......
Страница 18: ...General Information bkSpecto Ultrasound System 1 14 SM41406 4 This page is intentionally left blank...
Страница 94: ...Troubleshooting bkSpecto Ultrasound System 4 72 SM41406 4 6 Finish procedure online or by call Microsoft...
Страница 111: ...bkSpecto Ultrasound System Troubleshooting SM41406 4 4 89...
Страница 115: ...Adjustment Procedure bkSpecto Ultrasound System 5 4 SM41406 4 This page is intentionally left blank...
Страница 142: ...bkSpecto Ultrasound System Mechanical Parts SM41406 4 6 27 3 Remove the two speaker wires 4 Remove 11 M3X8 screws...
Страница 143: ...Mechanical Parts bkSpecto Ultrasound System 6 28 SM41406 4 5 Gently lift and slide out the Speaker Assembly...
Страница 149: ...Mechanical Parts bkSpecto Ultrasound System 6 34 SM41406 4 This page is intentionally left blank...
Страница 165: ...Theory of Operation bkSpecto Ultrasound System 8 12 SM41406 4 This page is intentionally left blank...