Troubleshooting
bkSpecto Ultrasound System
4-20
SM41406-4
Service Mode and Console mode
The scanner can be in one of two modes:
Console mode
and
Service mode
.
Console
mode
is the main mode which is used for clinical purposes where
Service mode
can be
accessed by BK Ultrasound certified technicians. Table 4-6 shows the differences between
Service mode
and
Console mode
. When the scanner is in
Service mode
File Explorer is
the shell and the user has access to the operation system. In
Console mode
the shell is the
BK Ultrasound Console with auto run and shortcuts being disabled.
In
Service mode
it is possible to open the
Console
application from the desktop and run it
with File Explorer as shell. Shut down the
Console
application using the
Service mode
button (ref. 4.18.1) or using an external keyboard (ref. 4.18.2).
CAUTION:
Always leave the scanner after technical service in
Console mode
. Only BK Ultrasound
certified staff is allowed to access the scanner in
Service mode
.
Service mode
Console Mode
Main
monitor
view
Shell:
Windows Explorer.
Console.
Users:
BK Ultrasound certified staff.
Clinical staff.
Switch
method:
Run
Console mode
application on
Desktop, 4.18.3.
Service mode button
and password.
OR
ALT + X
and password.
Table 4-6: Comparing scanner modes.
Entering Service Mode
1. Turn on the scanner and wait until the boot cycle is complete and boots in
Console
mode
.
2. Tap on the Tools button.
3. Tap on
SERVICE
.
4. Tap on SERVICE MODE.
5. Type in the Service Mode Password, chapter 4.28.
6. Press
OK
.
7. The scanner will restart and boot in
Service mode
.
Note:
In cases where this procedure does not work it is possible to boot from a bootable device. This
is described in section 4.26.
Содержание bkSpecto 1300
Страница 1: ...Service Manual Diagnostic Ultrasound System 1300 bkSpecto...
Страница 2: ......
Страница 4: ......
Страница 18: ...General Information bkSpecto Ultrasound System 1 14 SM41406 4 This page is intentionally left blank...
Страница 94: ...Troubleshooting bkSpecto Ultrasound System 4 72 SM41406 4 6 Finish procedure online or by call Microsoft...
Страница 111: ...bkSpecto Ultrasound System Troubleshooting SM41406 4 4 89...
Страница 115: ...Adjustment Procedure bkSpecto Ultrasound System 5 4 SM41406 4 This page is intentionally left blank...
Страница 142: ...bkSpecto Ultrasound System Mechanical Parts SM41406 4 6 27 3 Remove the two speaker wires 4 Remove 11 M3X8 screws...
Страница 143: ...Mechanical Parts bkSpecto Ultrasound System 6 28 SM41406 4 5 Gently lift and slide out the Speaker Assembly...
Страница 149: ...Mechanical Parts bkSpecto Ultrasound System 6 34 SM41406 4 This page is intentionally left blank...
Страница 165: ...Theory of Operation bkSpecto Ultrasound System 8 12 SM41406 4 This page is intentionally left blank...