Troubleshooting
bkSpecto Ultrasound System
4-24
SM41406-4
Replacement of Engine
1.
Follow the steps in
section
6 Mechanical Parts
for changing the engine.
2.
Disable encryption, ref. 4.40.3.2.
3.
Reprogram the serial number, ref. 4.23.
4. Import the user settings if available, ref. 4.47.2.
5. Setup DICOM if relevant, ref. 4.29.
6. Activate Windows if relevant, ref. 4.44.
7. Enable encryption if relevant, ref. 4.40.3.1.
8. Perform a safety test of the scanner according to
section 7 Preventative Maintenance
.
9. Check system performance using
section 3 Checking Procedure
.
10. Make sure the sure the system configuration is updated on the BK Ultrasound Service
Web.
Reprogramming the Serial Number
To Reprogram the Serial Number after Replacing the engine:
1. Ensure the system is running in Service Mode.
2. From the
Service Desktop
, select
Start Console
.
3. Follow 4.4 to change the
Windows time
and
Time Zone
if required.
4. Restart the system.
5. Follow 4.15 and take note of the system serial number (
License ID).
If it must be
changed, continue with the rest of the instructions. If no change is required, the task is
complete.
6. Exit the
Console
application by pressing ALT + X.
7. On the
Service Desktop
open
Janus Toolbox
.
8. Type
hwm
(1) and press
enter
.
Содержание bkSpecto 1300
Страница 1: ...Service Manual Diagnostic Ultrasound System 1300 bkSpecto...
Страница 2: ......
Страница 4: ......
Страница 18: ...General Information bkSpecto Ultrasound System 1 14 SM41406 4 This page is intentionally left blank...
Страница 94: ...Troubleshooting bkSpecto Ultrasound System 4 72 SM41406 4 6 Finish procedure online or by call Microsoft...
Страница 111: ...bkSpecto Ultrasound System Troubleshooting SM41406 4 4 89...
Страница 115: ...Adjustment Procedure bkSpecto Ultrasound System 5 4 SM41406 4 This page is intentionally left blank...
Страница 142: ...bkSpecto Ultrasound System Mechanical Parts SM41406 4 6 27 3 Remove the two speaker wires 4 Remove 11 M3X8 screws...
Страница 143: ...Mechanical Parts bkSpecto Ultrasound System 6 28 SM41406 4 5 Gently lift and slide out the Speaker Assembly...
Страница 149: ...Mechanical Parts bkSpecto Ultrasound System 6 34 SM41406 4 This page is intentionally left blank...
Страница 165: ...Theory of Operation bkSpecto Ultrasound System 8 12 SM41406 4 This page is intentionally left blank...