bkSpecto Ultrasound System
Mechanical Parts
SM41406-4
6-15
5. Remove the seven screws
6. It is now possible to remove the charger board assembly.
Remove the Main Mounting Plate
Initial tasks in chronological order:
6.2.4 Remove IO – Board
6.2.6 Remove Charger Board Assembly
Procedure
:
1. Check all the fan wires are labelled with the fan number. If a label is missing add a new
label.
2. Disconnect the six fan wires from the callisto main board. Labels on the wires indicate
the fan number.
Caution:
Remember to insert the wires in the order as shown above.
M3X6
M3X20
Wires from 4 & 5
Wires from 0,1,2 & 6
Metal
Brackets
Содержание bkSpecto 1300
Страница 1: ...Service Manual Diagnostic Ultrasound System 1300 bkSpecto...
Страница 2: ......
Страница 4: ......
Страница 18: ...General Information bkSpecto Ultrasound System 1 14 SM41406 4 This page is intentionally left blank...
Страница 94: ...Troubleshooting bkSpecto Ultrasound System 4 72 SM41406 4 6 Finish procedure online or by call Microsoft...
Страница 111: ...bkSpecto Ultrasound System Troubleshooting SM41406 4 4 89...
Страница 115: ...Adjustment Procedure bkSpecto Ultrasound System 5 4 SM41406 4 This page is intentionally left blank...
Страница 142: ...bkSpecto Ultrasound System Mechanical Parts SM41406 4 6 27 3 Remove the two speaker wires 4 Remove 11 M3X8 screws...
Страница 143: ...Mechanical Parts bkSpecto Ultrasound System 6 28 SM41406 4 5 Gently lift and slide out the Speaker Assembly...
Страница 149: ...Mechanical Parts bkSpecto Ultrasound System 6 34 SM41406 4 This page is intentionally left blank...
Страница 165: ...Theory of Operation bkSpecto Ultrasound System 8 12 SM41406 4 This page is intentionally left blank...