Preventative Maintenance
bkSpecto Ultrasound System
7-2
SM41406-4
7
Introduction
The purpose of preventative maintenance (PM) is to ensure performance and stability over
the lifetime of a system.
Note:
A PM should be performed once a year.
Preparing for a PM visit is essential as in most cases it is completed as a part of an
institution’s Continuity Agreement.
PM consists of eight steps. Some of these steps refer to specific sections in this
Service
Manual
. If an error occurs during a procedure, refer to section 4:
Troubleshooting
.
7.1
Required Equipment
The Equipment required for a PM visit can be divided into three sections:
Parts Required to Ensure System Performance and Stability
Check the Service Information bulletins (Hotline) posted on the BK Ultrasound Service Web
for any system modifications or important software updates to be completed.
Parts Required to Respond to Customer Requests and/or Complaints
Before arranging the PM visit, check to see if the customer has made any complaints or
requests. Be sure to bring the spare parts, software, and tools necessary to resolve any
issues.
Standard Tools and Equipment Required to Perform the PM
The equipment required to perform a PM is the same as used in section 3
Checking
Procedure
and
Electrical Safety Test
.
Section 3,
Checking Procedure
:
A Transducer.
USB flash key for exporting data.
A phantom for ultrasound.
Electrical Safety Test
:
Medical approved safety tester.
bkSpecto Electrical Safety Test Record (document number
TS43494
).
Setup for electrical safety test according to IEC/EN 60601-1 or IEC/EN 62353:2014.
Note:
As a precaution, BK Ultrasound recommends bringing the
Static Control Service Kit
in case
one of the static-sensitive boards must be accessed.
7.2
Performing a PM
Perform the procedures detailed in the following sections (7.2.1 to 7.2.10).
User Comments and Corrections
Check with the customer to ensure the system is operating in a satisfactory manner.
Содержание bkSpecto 1300
Страница 1: ...Service Manual Diagnostic Ultrasound System 1300 bkSpecto...
Страница 2: ......
Страница 4: ......
Страница 18: ...General Information bkSpecto Ultrasound System 1 14 SM41406 4 This page is intentionally left blank...
Страница 94: ...Troubleshooting bkSpecto Ultrasound System 4 72 SM41406 4 6 Finish procedure online or by call Microsoft...
Страница 111: ...bkSpecto Ultrasound System Troubleshooting SM41406 4 4 89...
Страница 115: ...Adjustment Procedure bkSpecto Ultrasound System 5 4 SM41406 4 This page is intentionally left blank...
Страница 142: ...bkSpecto Ultrasound System Mechanical Parts SM41406 4 6 27 3 Remove the two speaker wires 4 Remove 11 M3X8 screws...
Страница 143: ...Mechanical Parts bkSpecto Ultrasound System 6 28 SM41406 4 5 Gently lift and slide out the Speaker Assembly...
Страница 149: ...Mechanical Parts bkSpecto Ultrasound System 6 34 SM41406 4 This page is intentionally left blank...
Страница 165: ...Theory of Operation bkSpecto Ultrasound System 8 12 SM41406 4 This page is intentionally left blank...