83
USE AND MAINTENANCE MANUAL
7.2
Error Messages
PROBLEM
CAUSE
REMEDY
Overload
The scale is
overloaded.
Reduce the weight to a value
below the maximum scale
capacity.
Underload
The balance is
negative at least 20
divisions.
If on completion of the operation
the display shows a value other
than zero , press
to zero-
set the reading.
If the display is not zero-set on
pressing the key:
1) 1) switch off the terminal and
then switch it on again;
2) If the problem persists,
contact your Service Centre.
-01-
Converter fault.
Contact Assistance Service.
Scale connector
disconnected or
broken.
Switch off the terminal and
check that the connector is
properly connected.
If necessary, disconnect and
then re-connect the connector.
Switch the terminal back on.
Digital cells fail
to respond
Switch the terminal off and
then on again.
-06-
Serial number
error on digital
cell
Switch the terminal off and
then on again.
-07-
Digital cell serial
number error in
system with
several load
cells
Switch the terminal off and
then on again.
-08-
Digital cell serial
number error
Switch the terminal off and
then on again.
-09-
Digital cell serial
number and
configuration
error
Switch the terminal off and
then on again.
-10-
At least one cell
not configured
Switch the terminal off and
then on again.
-11-
Digital cell
powering error
Switch the terminal off and
then on again.
-12-
Powering error
in one digital cell
Switch the terminal off and
then on again.
-13-
Cell with internal
temperature
off-limits (-40 -
100) °C
Switch the terminal off and
then on again.
-21-
occasional
disconnection
temporary error
Change the battery
Internal lithium
battery
discharged
Contact Assistance Service.