Figure 45: CritiCool MINI System Message Troubleshooting Guide
Message
Cause of Problem
Action to
be taken
Comments
Connect Core Temperature Probe
No core temperature
probe is inserted in its
socket.
Connect core
temperature probe
* Pressing alarm mute
silences the buzzer for 30
minutes.
Check Water Tubes
Wrap is blocked due to
improper wrapping.
Wrap clamps are
closed.
Check for
creases, folds, or
objects that
obstruct the
water flow in the
wrap.
Check clamps.
* Pressing Alarm Mute
silences the buzzer for 30
minutes.
Check Core Temperature Probe
Misplacement of core
temperature probe in
core socket.
Core temperature
probe’s adapter is
connected to CritiCool
®
MINI without the
temperature probe.
Connect the core
temperature probe
to the appropriate
socket.
Connect
disposable
temperature probe
to the adapter and
insert into the
patient.
This alarm cannot be
muted.
Low Battery, Connect to Power
Supply
The battery of CritiCool
®
MINI has no power.
CritiCool
®
MINI
should be
attached to a
power supply.
This alarm cannot be muted.