Operation Panel Messages and Alerts
If the wrap’s tubes are connected, the temperature probes are connected, and
core temperature is measured, water circulation will start without additional user
action. If any of the above conditions is not fulfilled, the operation panel
message area displays technical and/or clinical alarm messages with a
sign.
NOTE:
Clinical alarms represent medium priority alarms while technical
messages represent lower priority alarms.
NOTE:
Sound pressure of the alarms is 67.5 dBA at a distance of 10
centimeters.
Constant alarms occur in the following states:
•
Halt condition
•
Select mode screen
The following messages should be checked and confirmed:
•
Low core temperature thermoregulation is continuing
•
Core readout too low
•
Out of normothermia range
•
Patient temperature above XX.XºC (*)
•
Patient temperature below YY.YºC (*)
•
Water temp too high (*)
NOTE:
Only authorized users can change the range of the alarms marked
by (*) in the settings screen. The user needs to insert a password to
enter the settings panel and change the alarm limit.
Figure 30: Adjustable Alarm Limits