DDT-320-000 Rev G Belmont Medical Technologies
Page
88
of
114
Figure 44: CritiCool MINI System Message Troubleshooting Guide
Message
Cause of Problem
Action to
be taken
Comments
Indicates that an alarm is activated
Tank is empty
No water in the tank.
Water tank float is
jammed
Open water tank
cap.
Refill water tank
to the maximum.
Insert a long
object to release
the float.
Add Water
Water level is too low
Refill water tank to
the maximum.
The alarm can be muted for
an unlimited time.
Connect Water Tubes
Connecting tubes are
not connected.
Connect
connecting tubes.
Check for
creases, folds, or
objects that
obstruct the water
flow in the wrap.
Check clamps.
* Pressing alarm mute
silences the buzzer for 30
minutes.