and operate as instructed.
NOTE:
When there is a difference between the Set-Point Temperature and
the Core Temperature, a further decrease in the Set-Point
temperature does not affect the water temperature in the wrap
.
NOTE:
Short transient changes in core temperature do not affect
thermoregulation and are compensated for by the system.
NOTE:
The rate of temperature change depends on several clinical
factors, including the size of the patient, medications being
administered, and health indicators.
Controlled Rewarming Mode
This mode is used for Controlled Rewarming following TTM.
The Controlled Rewarming mode enables warming the patient gradually
according to pre-configured rewarming steps.
The rewarming step rate is configured in the Settings Screen as shown
below. Available rates are offered from 0.05°C to 0.5°C per hour in
increments of 0.05°C.
The selected rewarming step is displayed on the main operating screen as
T/h
.
Figure 13: Selecting a Rewarming Step