Inserting and Attaching Temperature Probes
WARNING!!!
For proper use of the CritiCool
®
MINI device, the core
temperature probe must be inserted, and the surface
temperature probe must be attached to the patient per the
probes instructions for use. The location of the surface
temperature probe is a clinical decision. All temperature
probes directly measure temperature.
1. Insert the core temperature probe or gray adaptor cable (reusable or
disposable) into the left socket labeled “CORE” color-coded with gray on the
front of the device. (See
Front View
on page 18).
2. Insert the core temperature probe (reusable or disposable) into the
patient’s rectum or esophagus.
3. Insert the surface temperature probe or green adaptor cable (reusable or
disposable) into the right socket labeled “SURFACE” color-coded with green on
the front of the device.
4. Attach the surface temperature probe (reusable or disposable) to an exposed
area of skin with adhesive tape. When the patient is wrapped, the surface
temperature probe should not be under the CureWrap or covered.
NOTES:
•
The CritiCool
®
MINI device does not initiate thermoregulation if the Core
temperature probe is not properly positioned into the patient. Ensure that
direct patient feedback is monitored at all times.
•
The disposable temperature probes need to be connected to an adapter.
Make sure to connect the appropriate probe to its adapter (note the
labeling on the adapter).
•
Be sure to read and follow the instructions for use noted on the temperature
probe being used, paying particular attention to indications and
contraindications.