DDT-320-000 Rev G Belmont Medical Technologies
Page
57
of
114
Empty
This selection allows for emptying the system of the remaining water, prior
to storage of the CritiCool
®
MINI. This is recommended between cases.
To empty the water tank:
1. Switch to Standby Mode.
2. Disconnect the wrap from the system. Dispose of the wrap.
3. Connect a male draining connector to the "water out" of the CritiCool
®
MINI and direct the tube to a sink or 2-liter container for water
collection.
4. Touch the
MENU
icon.
5. Touch
Services
.
Touch
Empty
. The following screen appears.
6. When you are ready for the process to begin, touch
Start
. The
following screen appears.
Figure 28: Start Emptying Panel